Appendix C
70
The Zero Module can only be used in a program log mode. The Zero module
function is controlled through these two program mode options:
Auto Zero Interval
Interval between re-zeroing the instrument using
the Auto-Zero accessory.
Use Auto Zero
Select
Yes
to use the Zero Module. Select
No
to
not use the Zero Module.
Important points on Zero Module operation:
The Zero module will take 1 minute to take a zero reading. The first 45
seconds of that period is used to clear the chamber of particles.
Readings from last 15 second of the period, when the chamber is
cleared of particles, will be averaged to determine the Zero offset.
The log interval, when the Zero module is activated, must be 2 minutes
or greater. Data will not be recorded to the log file when the Zero module
is activated.