2: Setup and Operation
31
To initiate Waveform logging:
Press the
Print/Save
button on the Certifier test system keypad.
Select
Waveform
from the
Logging Features
section.
You may choose to
Add Comments
if you would like to annotate the data.
Click
OK
, add comments (if selected) and name the file.
The Certifier test system will collect 15 seconds of flowmeter waveform data and
then save it into the file.
Access the data by connecting the Certifier test system to a host computer via USB, or
by removing the SD Flash card from the Certifier test system and inserting it into a card
reader attached to a host computer.
2.3.9.2
Continuous Logging
The Continuous Logging feature will log snapshots of the same data being displayed on
the Certifier Parameter Screen and log it into a file on an SD Flash card. The logging
rate is once per second. Only data displayed on the Certifier Parameter Screen will be
logged. See section 2.3 on how to add or remove parameters. The data is exported into
“comma separated value” (CSV) format, readable by many common spreadsheet and
database applications. See
for selecting file delimeters other
than commas.
The Continuous logging feature requires an SD Flash card.
To initiate Continuous logging:
Press the
Print/Save
button on the Certifier test system keypad.
Select
Continuous
from the
Logging Features
section.
You may choose to
Add Comments
if you would like to annotate the data.
Click
OK
, add comments (if selected) and name the file.
The Certifier test system will continue to log data until the
Print/Save
button is
pressed again or until the SD Flash card fills up.
Access the data by connecting the Certifier test system to a host computer via USB, or
by removing the SD Flash card from the Certifier test system and inserting it into a card
reader attached to a host computer.
Note
: If the pause button was pressed prior to continuous logging, a slight delay may
be noticed in the start of the data acquisition.
Содержание Certifier FA Plus
Страница 1: ...CERTIFIER FA PLUS TEST SYSTEM OPERATOR S MANUAL P N 1980560 REVISION J FEBRUARY 2016...
Страница 2: ......
Страница 4: ......
Страница 10: ...vi Certifier FA Plus Test System This page intentionally left blank...
Страница 17: ...1 Introduction 7 Figure 5 Interface Module Touch Screen Display Key Pad Stylus...
Страница 20: ...Certifier FA Plus Test System 10 Figure 9 Oxygen Sensor Kit...
Страница 33: ...2 Setup and Operation 23 Figure 21 Parameter Definitions I E t E t I I E IP t E t I P f 1 t I P t E MV V TI f...
Страница 50: ...40 Certifier FA Plus Test System This page intentionally left blank...
Страница 58: ...48 Certifier FA Plus Test System This page intentionally left blank...