![TSI Incorporated Certifier FA Plus Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/tsi-incorporated/certifier-fa-plus/certifier-fa-plus_operators-manual_1178091039.webp)
2: Setup and Operation
29
2.3.9
Print/Save Button
Figure 26. Print/Save Options Screen
If the
Print/Save
button is pressed while viewing the Parameter Screen, the "Print/Save
Options" screen will appear. By selecting the various options on this screen you will be
able to:
Print the current data record without saving to a file.
Print the current data record and also save it to a file in a single operation.
Save the current data record to a file without printing.
Select
Waveform
logging function. This feature saves 15 seconds of "raw"
flowmeter data at a rate of 1 millisecond/reading to a file (requires an SD Flash
memory card).
Select
Continuous
logging function. This mode logs the Parameter Screen data
into a file at intervals of approximately 1 second until you command the logging to
stop (requires an SD Flash memory card).
When printing or saving, you can add custom comments to your data by selecting the
Add Comments
option. When that option is selected, an on-screen keyboard appears,
allowing you to enter your comment text.
Содержание Certifier FA Plus
Страница 1: ...CERTIFIER FA PLUS TEST SYSTEM OPERATOR S MANUAL P N 1980560 REVISION J FEBRUARY 2016...
Страница 2: ......
Страница 4: ......
Страница 10: ...vi Certifier FA Plus Test System This page intentionally left blank...
Страница 17: ...1 Introduction 7 Figure 5 Interface Module Touch Screen Display Key Pad Stylus...
Страница 20: ...Certifier FA Plus Test System 10 Figure 9 Oxygen Sensor Kit...
Страница 33: ...2 Setup and Operation 23 Figure 21 Parameter Definitions I E t E t I I E IP t E t I P f 1 t I P t E MV V TI f...
Страница 50: ...40 Certifier FA Plus Test System This page intentionally left blank...
Страница 58: ...48 Certifier FA Plus Test System This page intentionally left blank...