26
Certifier
®
FA Plus Test System
2.3.6
Breath Trigger Types
This screen defines how the start of the inspiratory breath cycle and the expiratory
breath cycle are detected. Under most circumstances, it is recommended that the auto-
trigger be used.
Figure 24. Trigger Options Menu
Flow Rate
Start and end of the breath is determined by the specified
flow rate. If Auto-Triggering is selected, instrument tries to
determine trigger levels. In some cases you may need to
manually select the flow rates by looking at the flow
waveform on the Graph Screen.
Pressure
Start and end of the breath is determined by the specified
pressure levels. Positive slope changed through the first
value is used for the start of inspiratory and a negative
slope through the second value is used for the start of
expiratory.
Note
: This trigger type is intended to give
advance users additional setup options for special
circumstances, such as achieving basic breath
measurements from high frequency ventilators or other
setups in which flow rate or auto triggering are non-ideal.
TTL
Start and end of the breath is determined by a TTL voltage
signal given at the connector on the High Flow Module
labeled “Trigger Input”. This is only available for the 4081
High Flow Module. The connector is a 3.5 mm mono audio
jack plug.
Содержание Certifier FA Plus
Страница 1: ...CERTIFIER FA PLUS TEST SYSTEM OPERATOR S MANUAL P N 1980560 REVISION J FEBRUARY 2016...
Страница 2: ......
Страница 4: ......
Страница 10: ...vi Certifier FA Plus Test System This page intentionally left blank...
Страница 17: ...1 Introduction 7 Figure 5 Interface Module Touch Screen Display Key Pad Stylus...
Страница 20: ...Certifier FA Plus Test System 10 Figure 9 Oxygen Sensor Kit...
Страница 33: ...2 Setup and Operation 23 Figure 21 Parameter Definitions I E t E t I I E IP t E t I P f 1 t I P t E MV V TI f...
Страница 50: ...40 Certifier FA Plus Test System This page intentionally left blank...
Страница 58: ...48 Certifier FA Plus Test System This page intentionally left blank...