Chapter 4
24
Note
: DIP switch 3 or 4 must be set to the “User Calibration” (ON)
position or the Q-T
RAK
will not go into calibration mode. Please
refer to Appendix B: “Internal DIP Switch Settings.”
3. Compare the reading on the Q-T
RAK
with the reading from the
reference device. If they differ, use the arrow keys (
!"
) on the
Q-T
RAK
to adjust the Q-T
RAK
display as needed so that the displayed
measurement matches the measurement indicated by the reference
device. Press the arrow key briefly to change the display 0.1 degree or
0.1% at a time. Hold the arrow key down to change more quickly.
Note:
If the two readings are the same, there is no need to proceed with the
calibration. Press the
CO
2
,
TEMP,
or
HUMIDITY
key to terminate
the calibration procedure.
4. Press
the
SAMPLE
key to take a measurement. The Q-T
RAK
will
display a 15-second count-down. When the count-down is completed,
the Q-T
RAK
will go back to survey mode.
5. Observe the displayed reading on the Q-T
RAK
, it should agree with the
reference device within specifications (see Appendix A). If not, repeat
the calibration.
6. If the displayed reading is accurate, the calibration is completed.
Calibrating the CO Concentration Measurement (Model 8551
only)
TSI recommends calibrating the Q-T
RAK
CO measurement monthly to help
ensure accurate readings. The CO concentration measurement is affected by
changes in temperature and atmospheric pressure. Normal day-to-day
variations due to local weather conditions have little effect. However,
changes in altitude can cause more significant errors. For best accuracy,
calibrate the Q-T
RAK
CO measurement for your local conditions or if
conditions change.
To calibrate the CO sensor, follow the procedure below. You will need a
cylinder of pure air for the zero calibration and a cylinder of gas with a
known concentration of CO for the span calibration, 35 ppm or 200 ppm is
recommended. Contact your local TSI distributor for available CO
calibration kits.
Note
: Calibration with CO gas should always be done in a well-ventilated
area.
If necessary, you can select between the factory CO calibration and a user
calibration. Please refer to Appendix B: “Internal DIP Switch Settings.”