![TSI Incorporated 8550 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/tsi-incorporated/8550/8550_operation-and-service-manual_1178093016.webp)
Chapter 3
10
Programming the Available Time Constants
To program the list of time constants using Q-T
RAK
Data Analysis Software:
1. Make sure the Q-T
RAK
is connected to the computer and turned on.
2.
Select
Parameters,
then
Time Constants
from the
Instrument Setup
menu. T
RAK
P
RO
retrieves the current time constant settings from the
Q-T
RAK
and displays them in the following dialog:
3.
Enter a value for each of the five available time constants. (The range is
limited to 2–60 seconds.)
4.
Select
Send
.
The Q-T
RAK
is reprogrammed to offer the time constants you have
specified.
STATISTICS Key
Use the
STATISTICS
key to sequentially view the average, minimum, and
maximum readings as well as the elapsed time of the most recently sampled
data. You may view sample statistics while the sample is in-progress;
however, one time-constant must have elapsed first. If one of the LOG
modes is active, a test identification number is displayed also. Press the
STATISTICS
key once to display the average reading, again to display the
minimum reading, again to display the maximum reading, and again to
display the elapsed time for that sample (and again for the test ID if in LOG
mode). If you press the
STATISTICS
key a fifth time (sixth time if in LOG
mode), the Q-T
RAK
switches back into the currently selected measuring
mode. You must sequence through all four statistic displays (i.e., press the
STATISTICS
key five times, six if in LOG mode) before the Q-T
RAK
goes
back into the currently selected measuring mode. All parameters are