The Model 8495 T
RANSDUCER
D
ISPLAY
& M
ONITOR
displays an output that
is linearly proportional to the input signal.
When the T
RANSDUCER
D
ISPLAY
& M
ONITOR
is first turned on it will scroll
through the current settings for the signal input, full scale, and time constant.
The T
RANSDUCER
D
ISPLAY
& M
ONITOR
comes with the parameters set to
the default values found in the following table.
Factory Settings
Time Constant
1
Full Scale
10000
Signal Input
0-5 V
Zero Adjust
0
Span Adjust
1.000
Low Alarm
Off
High Alarm
Off
LEDs
Unlatched
Relays
Unlatched
Audible Alarm
Off
C
ERTIFIER
®
Off
The
SET
key is used to change (or view) the settings of the above
parameters. By pressing the
SET
key repeatedly, the parameters can be
changed (or viewed). Each time the up and down arrow keys (
) are used
to change a value the
SET
key must be pressed to advance to the next
parameter.
•
Full scale
The full scale can be adjusted by pressing the up and down arrow keys
(
). The available values are: 0.1000 to 99990.
•
Signal input
The signal input can be adjusted by pressing the up and down arrow
keys (
). The available choices are: 0-5 VDC, 0-10 VDC,
1-5 VDC, 2-10 VDC, 0-20 mA, 4-20 mA. For 0-20 mA or 4-20 mA
settings, use the provided jumper to short the MA SIG+ terminal to the
5V SIG+ terminal.