Chapter 2 

 
 

Setting Up and Operating the Model 8026 
Particle Generator 

Introduction 

The Model 8026 Particle Generator is intended to supplement the 
naturally occurring particles in the air when performing quantitative 
respirator fit tests using the TSI PortaCount

®

 Respirator Fit Tester. The 

following setup and operating instructions presume this intended usage.  
 
This does not preclude using the Model 8026 Particle Generator for 
other applications to generate a moderate concentration of poly-disperse 
particles. Some general notes on alternate uses for the Model 8026 are 
found at the end of this chapter. 
 
Before setting up the Model 8026 Particle Generator, it is important to 
select a suitable location for fit testing. 

Selecting a Fit Testing Location 

The particle generator may be used with the PortaCount fit tester if 
you are unable to obtain the minimum particle concentration. 
 

I m p o r t a n t   C o n s i d e r a t i o n s  

 

The Particle Generator or any other sources of particle generation 
should be kept at least 6 feet from the PortaCount Pro fit tester during 
operations.

 

 

For the Particle Generator to function properly, operate the generator 
and conduct the fit tests in an enclosed area. A room smaller than 
400 sq. ft. would provide the best conditions. Particle generation will not 
function as efficiently in an open cubicle area or a very large room. 

 

The Particle Generator and other forms of particle generation should 
only be used sparingly, when needed. Often they are only needed in the 
morning hours or at the beginning of fit testing. Do 

not

 operate any form 

of particle generation if ambient particle concentrations are above 
8,000 pt/cc (as per use with full- & half-mask, P100 fit testing) or 
800 pt/cc (as per use with N95 fit testing).

 

 

Never split the Twin Tubing assembly in order to move the ambient 
(blue) tube closer to the Particle Generator.

 

Содержание 8026

Страница 1: ...PARTICLE GENERATOR MODEL 8026 OPERATION AND SERVICE MANUAL P N 1980319 REVISION M MARCH 2016 ...

Страница 2: ......

Страница 3: ...nstrument is opened by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning g This warranty is VOID if the product has been misused neglected subjected to accidental or intentional damage or is not properly installed maintained or cleaned according to the requ...

Страница 4: ...TATION OF WARRANTY AND LIABILITY which contains the complete and exclusive limited warranty of Seller This LIMITATION OF WARRANTY AND LIABILITY may not be amended modified or its terms waived except by writing signed by an Officer of Seller Service Policy Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers our service policy is designed to give ...

Страница 5: ...g a Fit Testing Location 3 Setting up the Model 8026 Particle Generator 4 Operating the Model 8026 Particle Generator 6 3 MAINTENANCE 7 Routine Maintenance 7 Cleaning the Atomizer Nozzle 7 4 TROUBLESHOOTING 9 Appendix A SPECIFICATIONS 11 AVAILABLE APPLICATION NOTES To obtain any of the Application Notes contact TSI at U S 800 874 2811 651 490 2811 Fax 651 490 3824 International 001 651 490 2811 Fa...

Страница 6: ...FETY INFORMATION The Model 8026 Particle Generator includes a high temperature safety switch If the generator overheats this switch shuts off the unit permanently and it must be returned to TSI for servicing ...

Страница 7: ...missing or damaged Table 1 1 Model 8026 Particle Generator Component Parts Quantity Item Description Part Model 1 Particle generator 8026 1 Particle generator power cord 1303567 1 Package of 100 salt tablets 803011 1 Model 8026 Particle Generator Operation and Service Manual 1980319 Parts Identification for the Particle Generator Figure 1 1 identifies the parts of the Model 8026 Particle Generator...

Страница 8: ... replacement parts and supplies for the Model 8026 Particle Generator Table 1 2 Replacement Parts and Supplies Quantity Item Description Part Model 1 Package of 100 salt tablets 803011 1 Storage lid for reservoir 2002030 1 Power cord North America 1303567 1 Atomizer assembly 2401041 ...

Страница 9: ...ster if you are unable to obtain the minimum particle concentration Important Considerations The Particle Generator or any other sources of particle generation should be kept at least 6 feet from the PortaCount Pro fit tester during operations For the Particle Generator to function properly operate the generator and conduct the fit tests in an enclosed area A room smaller than 400 sq ft would prov...

Страница 10: ...it more quickly 3 Place the temporary cover tightly on the reservoir and shake it gently to dissolve the salt tablet 4 Once the solution is mixed remove the temporary cover and screw the reservoir onto the particle generator See Figure 2 1 5 Place the particle generator on the floor in a corner of the testing room Make sure the ventilation louvers on the back and bottom of the generator are not ob...

Страница 11: ... to stabilize The generator must operate in an upright position at all times One reservoir full of salt solution will last from 8 to 16 hours depending on the generator setting The salt solution can be saved and re used Do not mix old solution with fresh solution To mix new solution first discard whatever remains in the reservoir To re use the solution detach the reservoir from the generator cover...

Страница 12: ... Screw Increase Figure 2 3 To restore the output to the factory setting turn the screw completely clockwise to maximize the aerosol flow Then make three 3 complete counterclockwise turns The screw should turn easily Do not force it Alternate Uses for the Particle Generator The Model 8026 Particle Generator may be used as a generalized source of poly disperse particles in an environment When used f...

Страница 13: ...Periodically wipe down the outside surface of the generator using a damp cloth After each use shut off the generator remove the salt solution reservoir and then switch the generator back on for 40 seconds This will purge salt solution from the unit and reduce contamination of the carrying case No other routine periodic maintenance is required If a decrease in output is noticed clean the atomizer n...

Страница 14: ...compressed air to dry the parts or air dry them Note The orifice is easily damaged Do not insert anything in the orifice in an attempt to clean it 4 Reassemble and install the atomizer jet making sure to replace the O ring When replacing the retainer make sure it is pushed in completely 5 Attach the draw tube 6 Test the generator for proper operation output using clean tap water ...

Страница 15: ...instructions in Chapter 2 Mix a new batch of solution and fill the reservoir Clean the atomizer head Refer to the maintenance procedure in the previous chapter Adjust the pressure to a higher setting as necessary Refer to the procedure in Chapter 2 Add 1 or 2 additional tablets of salt to mixture Or move particle generator closer to PortaCount fit tester Do NOT locate closer than 4 ft from PortaCo...

Страница 16: ...Chapter 4 10 This page intentionally left blank ...

Страница 17: ... Size 10 in x 4 in x 5 25 in 25 4 cm x 10 1 cm x 13 3 cm Weight 2 75 lb 1 25 kg Generated Particle Statistics using 2 NaCl solution in distilled water Count median diameter CMD Geometric standard deviation GSD 0 04 micrometers nominal 2 2 nominal Operating Voltage 115 or 230 VAC check label on unit Warranty One year on workmanship and materials ...

Страница 18: ......

Страница 19: ...e information USA Tel 1 800 874 2811 UK Tel 44 149 4 459200 France Tel 33 1 41 19 21 99 Germany Tel 49 241 523030 India Tel 91 80 67877200 China Tel 86 10 8219 7688 Singapore Tel 65 6595 6388 P N 1980319 Rev M 2016 TSI Incorporated Printed in U S A 1980319 ...

Отзывы: