Troubleshooting
This chapter describes the symptoms, possible causes and
recommended solutions for common problems encountered with the
Model 8026 Particle Generator.
Some common problems that could be encountered include:
Symptom
Possible Cause
Corrective Action
Low particle
counts
The particle generator is
not running.
The salt solution is mixed
incorrectly.
The salt solution reservoir
is low or empty.
The atomizer head on the
particle generator is
clogged.
The particle generator
pressure is set too low.
Salt solution may not be
suitable for room.
Turn on the power to the
particle generator.
Refer to the mixing
instructions in Chapter 2.
Mix a new batch of solution
and fill the reservoir.
Clean the atomizer head.
Refer to the maintenance
procedure in the previous
chapter.
Adjust the pressure to a
higher setting, as
necessary. Refer to the
procedure in Chapter 2.
Add 1 or 2 additional tablets
of salt to mixture. Or move
particle generator closer to
PortaCount fit tester. Do
NOT
locate closer than 4 ft.
from PortaCount fit tester.
Low particle
counts
(cont.)
The testing room is
inappropriate.
It is too large
It is not a closed room
It has too high an air
exchange rate
If the room is too large or
not closed, select another
location. If the air exchange
rate is too high, increase
the output of the particle
generator.
No aerosol
output
The compressor does not
start when turned on. (Red
light comes on okay).
Return the generator to TSI
for servicing.