![TSI Incorporated 8026 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/tsi-incorporated/8026/8026_operation-and-service-manual_1178063010.webp)
Chapter 2
4
Setting up the Model 8026 Particle Generator
The particle generator supplements the naturally occurring particles
in the room environment. Follow the steps below to set up the particle
generator:
1. Fill the reservoir jar with clean tap water to the fill line. Use room
temperature water to help the salt tablet dissolve more quickly.
Fill line
Figure 2-1
2. Drop one salt tablet into the reservoir. Each tablet contains
100 mg of salt. Break or crush the tablet before dropping it into
the water to dissolve it more quickly.
3. Place the temporary cover tightly on the reservoir and shake it
gently to dissolve the salt tablet.
4. Once the solution is mixed, remove the temporary cover and
screw the reservoir onto the particle generator. See Figure 2-1.
5. Place the particle generator on the floor in a corner of the testing
room. Make sure the ventilation louvers on the back and bottom
of the generator are not obstructed, as this could cause the
instrument to overheat.
C a u t i o n
Do
not
place the Particle Generator on the table or
benchtop. Vibrations from the internal compressor may
cause the generator to move across the table. If left
unattended, the generator could fall on the floor resulting
in damage to the instrument.