Setting Up and Operating the Model 8026 Particle Generator
5
6. Plug the female end of the power cord into the receptacle on the
back of the particle generator. Plug the male end into a grounded
wall outlet and turn on the power to the generator. The power switch
is located next to the power cord receptacle. See Figure 2-2.
Power
switch
Power cord
Figure 2-2
Run the particle generator in the testing room, with the door closed,
for at least 15 minutes before beginning fit testing to allow the particle
count to stabilize. The generator must operate in an upright position
at all times.
One reservoir full of salt solution will last from 8 to 16 hours,
depending on the generator setting. The salt solution can be saved
and re-used. Do
not
mix old solution with fresh solution. To mix new
solution, first discard whatever remains in the reservoir.
To re-use the solution, detach the reservoir from the generator, cover
it tightly with the temporary cover, and store it in its designated space
inside the carrying case.
C a u t i o n
Never attempt to store the particle generator in the
carrying case with a full or partially-filled reservoir
attached, as the solution will leak out and damage the
generator and/or carrying case.