Setting Up and Operating the Model 8026 Particle Generator 

6.  Plug the female end of the power cord into the receptacle on the 

back of the particle generator. Plug the male end into a grounded 
wall outlet and turn on the power to the generator. The power switch 
is located next to the power cord receptacle. See Figure 2-2. 

 

Power 
switch

Power cord

 

Figure 2-2 

 

Run the particle generator in the testing room, with the door closed, 
for at least 15 minutes before beginning fit testing to allow the particle 
count to stabilize. The generator must operate in an upright position 
at all times. 
 
One reservoir full of salt solution will last from 8 to 16 hours, 
depending on the generator setting. The salt solution can be saved 
and re-used. Do 

not

 mix old solution with fresh solution. To mix new 

solution, first discard whatever remains in the reservoir. 
 
To re-use the solution, detach the reservoir from the generator, cover 
it tightly with the temporary cover, and store it in its designated space 
inside the carrying case.  
 

 

C a u t i o n

 

 

Never attempt to store the particle generator in the 
carrying case with a full or partially-filled reservoir 
attached, as the solution will leak out and damage the 
generator and/or carrying case.

 

Содержание 8026

Страница 1: ...PARTICLE GENERATOR MODEL 8026 OPERATION AND SERVICE MANUAL P N 1980319 REVISION M MARCH 2016 ...

Страница 2: ......

Страница 3: ...nstrument is opened by anyone other than a factory authorized service center with the one exception where requirements set forth in the manual allow an operator to replace consumables or perform recommended cleaning g This warranty is VOID if the product has been misused neglected subjected to accidental or intentional damage or is not properly installed maintained or cleaned according to the requ...

Страница 4: ...TATION OF WARRANTY AND LIABILITY which contains the complete and exclusive limited warranty of Seller This LIMITATION OF WARRANTY AND LIABILITY may not be amended modified or its terms waived except by writing signed by an Officer of Seller Service Policy Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our customers our service policy is designed to give ...

Страница 5: ...g a Fit Testing Location 3 Setting up the Model 8026 Particle Generator 4 Operating the Model 8026 Particle Generator 6 3 MAINTENANCE 7 Routine Maintenance 7 Cleaning the Atomizer Nozzle 7 4 TROUBLESHOOTING 9 Appendix A SPECIFICATIONS 11 AVAILABLE APPLICATION NOTES To obtain any of the Application Notes contact TSI at U S 800 874 2811 651 490 2811 Fax 651 490 3824 International 001 651 490 2811 Fa...

Страница 6: ...FETY INFORMATION The Model 8026 Particle Generator includes a high temperature safety switch If the generator overheats this switch shuts off the unit permanently and it must be returned to TSI for servicing ...

Страница 7: ...missing or damaged Table 1 1 Model 8026 Particle Generator Component Parts Quantity Item Description Part Model 1 Particle generator 8026 1 Particle generator power cord 1303567 1 Package of 100 salt tablets 803011 1 Model 8026 Particle Generator Operation and Service Manual 1980319 Parts Identification for the Particle Generator Figure 1 1 identifies the parts of the Model 8026 Particle Generator...

Страница 8: ... replacement parts and supplies for the Model 8026 Particle Generator Table 1 2 Replacement Parts and Supplies Quantity Item Description Part Model 1 Package of 100 salt tablets 803011 1 Storage lid for reservoir 2002030 1 Power cord North America 1303567 1 Atomizer assembly 2401041 ...

Страница 9: ...ster if you are unable to obtain the minimum particle concentration Important Considerations The Particle Generator or any other sources of particle generation should be kept at least 6 feet from the PortaCount Pro fit tester during operations For the Particle Generator to function properly operate the generator and conduct the fit tests in an enclosed area A room smaller than 400 sq ft would prov...

Страница 10: ...it more quickly 3 Place the temporary cover tightly on the reservoir and shake it gently to dissolve the salt tablet 4 Once the solution is mixed remove the temporary cover and screw the reservoir onto the particle generator See Figure 2 1 5 Place the particle generator on the floor in a corner of the testing room Make sure the ventilation louvers on the back and bottom of the generator are not ob...

Страница 11: ... to stabilize The generator must operate in an upright position at all times One reservoir full of salt solution will last from 8 to 16 hours depending on the generator setting The salt solution can be saved and re used Do not mix old solution with fresh solution To mix new solution first discard whatever remains in the reservoir To re use the solution detach the reservoir from the generator cover...

Страница 12: ... Screw Increase Figure 2 3 To restore the output to the factory setting turn the screw completely clockwise to maximize the aerosol flow Then make three 3 complete counterclockwise turns The screw should turn easily Do not force it Alternate Uses for the Particle Generator The Model 8026 Particle Generator may be used as a generalized source of poly disperse particles in an environment When used f...

Страница 13: ...Periodically wipe down the outside surface of the generator using a damp cloth After each use shut off the generator remove the salt solution reservoir and then switch the generator back on for 40 seconds This will purge salt solution from the unit and reduce contamination of the carrying case No other routine periodic maintenance is required If a decrease in output is noticed clean the atomizer n...

Страница 14: ...compressed air to dry the parts or air dry them Note The orifice is easily damaged Do not insert anything in the orifice in an attempt to clean it 4 Reassemble and install the atomizer jet making sure to replace the O ring When replacing the retainer make sure it is pushed in completely 5 Attach the draw tube 6 Test the generator for proper operation output using clean tap water ...

Страница 15: ...instructions in Chapter 2 Mix a new batch of solution and fill the reservoir Clean the atomizer head Refer to the maintenance procedure in the previous chapter Adjust the pressure to a higher setting as necessary Refer to the procedure in Chapter 2 Add 1 or 2 additional tablets of salt to mixture Or move particle generator closer to PortaCount fit tester Do NOT locate closer than 4 ft from PortaCo...

Страница 16: ...Chapter 4 10 This page intentionally left blank ...

Страница 17: ... Size 10 in x 4 in x 5 25 in 25 4 cm x 10 1 cm x 13 3 cm Weight 2 75 lb 1 25 kg Generated Particle Statistics using 2 NaCl solution in distilled water Count median diameter CMD Geometric standard deviation GSD 0 04 micrometers nominal 2 2 nominal Operating Voltage 115 or 230 VAC check label on unit Warranty One year on workmanship and materials ...

Страница 18: ......

Страница 19: ...e information USA Tel 1 800 874 2811 UK Tel 44 149 4 459200 France Tel 33 1 41 19 21 99 Germany Tel 49 241 523030 India Tel 91 80 67877200 China Tel 86 10 8219 7688 Singapore Tel 65 6595 6388 P N 1980319 Rev M 2016 TSI Incorporated Printed in U S A 1980319 ...

Отзывы: