Core Modules
39
WARNING
Health hazard
Locations that are contaminated, e.g., by dust, cannot be effectively sterilized by all
sterilization routines. Thoroughly clean surfaces can be adequately sterilized only.
•
If there is a risk of inadequate sterilization, then you must continue to observe all the necessary
safety measures.
•
Dust deposits inside devices can result in ineffective sterilization.
DANGER
Hazardous voltage
Penetration of liquids will cause a short-circuit.
Keep liquids away from the ventilation openings.
This warning is given on instrument´s server rack housing, attached onto the front glass door.
NOTICE
- Instrument damage!
Process Control Units have openings in their housing. Boards inside Process Control Units
can be damaged by contact with liquids.
•
Use moistened wipes only, i.e. wipes that are not saturated or dripping, for cleaning of the
Process Control Units!
•
Never use disinfectant sprays.
NOTICE
- Instrument damage!
Sensors can be damaged by contact with H2O2.
Units with gas sensors (e.g. O2, CO2) must have their test air inlets and outlets sealed.
10.4. Cable
Organic solvents must not be used for cleaning purposes. Avoid the use of cleaning sprays, spray
disinfectants, or strong detergents, as they could damage the cables.
The cables can be wiped clean with a clean soft lint-free cloth, which can, if required, be moistened
with water or a pH-neutral detergent. Detergent residues should then be wiped off.
The cables can be decontaminated with H2O2 vapor
before they are brought into the test laboratory.
10.5. Rack
Racks should be wiped off, if required, with a moistened soft, lint-free cloth or tissue to remove any
adhering dirt. A stainless steel cleaning agent can also be used to remove any stubborn dirt.
Detergent residues should then be wiped off.
After cleaning, the shelves can be decontaminated with H2O2 vapor.