ID5000, XR5000 Level 1 Service Manual
Issue 4, 8/2017
29
5
Refit the rubber corner
bumpers (see page 23).
6
Check that the indicator operates and place on charge for four hours. The indicator must be off during charging to ensure
that the battery is fully charged.
Replacing the load bar and power/comms connectors
Figure 12: C16 power/comms
and load bar connectors
The procedure for replacing the load bar and power/comms connector assemblies is the same.
Removing the power/comms and load bar connectors
1
Remove the rubber corner bumpers (see page 23).
2
Remove the indicator case back (see page 25).
3
Unplug the battery from the PCA.
4
Slide the FPC of the connector out of the connector body on the PCA.
5
Remove the dust cap, unscrew the two screws securing the connector, then remove the assembly.
Refitting the power/comms and load bar connectors
1
Check that the case is clean where the connector meets the case and that all plastic swarf is removed.
The orientation of the FPC on the connector is facing up, out of the case.
2
Replace the dust cap.
3
Tighten the two screws using a torque-limited screwdriver set to 0.5 Nm (4 lbf-in), or hand screwdriver.
4
Gently slide the FPC connector into the connector body on the PCA. Ensure that the FPC connector is connected properly
(straight and the correct way around).
5
Plug the battery connector into the PCA.
6
Refit the case
back (see page 25).
7
Refit the rubber corner bumpers see page 23).
Removing and refitting the USB Dust Cap
Removing the USB dust cap
1
Remove the rubber corner bumpers (see page 23).
2
Remove the indicator case back (see page 25).
3
Remove the USB dust cap by pulling it off the USB port recess (Step 1) and lifting it out of the dovetail groove (Step 2).