6-18
If the number of calibration points is greater than
2
, the
oven continues to request each mass until all calibration
points are complete. The oven then displays:
deg of freedom = X
CC = X.XXXXX
SEE = X.XXXXX
Press Enter
where:
CC
=
Correlation coefficient, which indicates the
strength of the relationship between two
variables. In this case, the two variables are the
mass placed on the scale and the
corresponding response of the scale.
Correlation coefficients are always between –
1
and
+1
. A value close to
0
indicates a weak
relationship between the variables. If two
variables have a
negative
linear relationship
(one variable increases as the other decreases),
their correlation coefficient is close to
–1
. If
two variables have a
positive
linear
relationship (one variable increases as the
other increases), their value is close to
+1
. In an
ideal scale calibration, the
CC
value would be
close to
+1
.
SEE
= Standard error of estimate, which is a measure
of the linearity of the data used to calibrate the
scale. The more linear the scale calibration data,
the smaller the value of
SEE
. In an ideal scale
calibration, the
SEE
value would be less than
0.3
.
Press (ENTER). The oven then asks, "Do you want to
accept this calibration?".
Press(YES) to accept the calibration. The oven stores the
scale calibration and returns to the Scale Menu.
After a successful calibration, place a known mass (ANSI/ASTM
Class 4 or NIST Class P) on the hearth plate. Verify the scale as
described on page 6–15, then remove the known mass.
Содержание ICO 4740
Страница 2: ......
Страница 13: ...ICO xi Figure 21 Disengage Paper Spindle A 10 Figure 22 Insert New Paper Roll A 11...
Страница 14: ......
Страница 22: ...1 8 Figure 1 Parts and Accessories...
Страница 35: ...ICO 3 1 Chapter 3 INSTALLATION This chapter provides instructions on site selection and installation of the ICO...
Страница 102: ...5 30 and then requests the sample number as described in the Active Project Selected section on page 5 28...
Страница 138: ......
Страница 139: ...ICO A 1 Appendix A MAINTENANCE AND SERVICE This appendix contains information on maintaining and servicing the ICO...
Страница 152: ......
Страница 153: ...ICO B 1 Appendix B MENU MAP This appendix provides a map of the menus in the ICO control software...
Страница 161: ...ICO C 1 Appendix C SPECIFICATIONS This appendix contains specifications for the ICO...
Страница 166: ......
Страница 170: ......