General information regarding safety

Only skilled qualified personnel are allowed to 

perform the described work on the control panel.

Only skilled qualified electricians are allowed to

work on the electrical system.
For all work performed on EASYLAB components,

the following regulations and guidelines must be

complied with. This applies in particular to the 

following German country specific regulations or 

as appropriate in the country where the installation

is taking place:
• Equipment and Product Safety Laws (GPSG)
• Industrial Health and Safety Regulations

(BetrSichV)

• Accident Prevention Regulations (BGV A1, 

BGV A3)

General safety measures

• Large temperature differences

Condensation can damage the electronics 

beyond repair. If the control panel has been 

kept in an unheated area, wait at least two 

hours before switching on the supply voltage 

for commissioning.

• Electrostatic charge

Electrostatic charge can damage the electronics.

To avoid this, first touch an equipotentially 

bonded metal surface, e.g. a water pipe, for 

electrical earthing before you remove the control

panel from its protective wrapping.

• Fixing the base plate and the adapter

Tighten the screws only hand-tight in order to not

damage the base plate or the adapter.

• Foreign matter and liquids

If liquid gets inside the control panel, let 

the control panel completely dry before 

commissioning. Remove foreign matter, if any.
If the device emits a smell or smoke, have it 

checked by the manufacturer.

4

BE-LCD-01 Operating and Installation Manual / Leaflet no. M375EV8 (01/2015)

2 Safety and correct use

Correct use

The BE-LCD-01 control panel is an EASYLAB 

component intended for the display of values and

for performing various functions. If the control panel

is connected to a fume cupboard controller, the

function display complies with EN 14175; operating

functions and special functions can be set by 

the user. If a room is equipped with EASYLAB 

controllers, the control panel is intended for 

monitoring and for operating mode default setting.
• Connect the control panel to an EASYLAB TCU3

used for fume cupboard control (equipment

function FH-xxx).

• Connect the control panel to a EASYLAB TCU3

controller used for extract air control (equipment

function RE), supply air control (equipment

function RS), or room pressure control 

(equipment function PC).

• Connect the control panel to an EASYLAB TAM

adapter module with active room management

function (RMF).

• Surface mount the control panel onto the side

frame of the fume cupboard.

• As an alternative, mount the control panel to a

wall, with our without junction box.

• Observe the technical data for the control panel.

Incorrect use

Do not use the control panel outdoors, in wet areas,

or in potentially explosive atmospheres.

Residual risks

A failure of the supply voltage is indicated on the

control panel only if the EASYLAB controller is

equipped with the EM-TRF-USV expansion module

and the battery pack has been connected and fully

charged.

Содержание EASYLAB BE-LCD-01

Страница 1: ...ating and Installation Manual Leaflet no M375EV8 01 2015 The art of handling air The art of handling air Operating and Installation Manual EASYLAB Control panel BE LCD 01 for fume cupboard control and...

Страница 2: ...Fume cupboard lighting ________________ 10 Overriding operating mode defaults set by the central BMS____________________ 10 Operating states alarm messages fault displays ________________________ 11 5...

Страница 3: ...lation manual before starting any work and to comply with it The manual must be given to the facilities manager when handing over the system The facilities manager must include the manual with the sys...

Страница 4: ...reign matter if any If the device emits a smell or smoke have it checked by the manufacturer 4 BE LCD 01 Operating and Installation Manual Leaflet no M375EV8 01 2015 2 Safety and correct use Correct u...

Страница 5: ...green yellow red with text HIGH and LOW Connection socket for service and commissioning device Acoustic alarm acknowledgement Manual control Selection of operating mode Lighting or other equipment RM...

Страница 6: ...tion One button for the fume cupboard lighting One button for manual control Service socket The range of functions of the control panel can be adapted to project specific requirements using the EasyCo...

Страница 7: ...udes BE LCD 01 control panel with adapter Connecting cable approx 5 m patch cable blue Operating and installation manual Transport on site If possible take the control panel in the transport packaging...

Страница 8: ...of fume cupboard operating states and faults Display texts p 11 onwards The display of additional information can be configured Volume flow rate setpoint and actual values Face velocity actual value o...

Страница 9: ...me flow will then automatically be deactivated after the set delay Additional functions Depending on the configuration and operating mode default setting additional displays and function buttons may b...

Страница 10: ...ny other operating mode on the control panel This operating mode default setting can be terminated by pressing the manual control button again The fume cupboard controller then resumes the previous op...

Страница 11: ...e controller configuration the damper blade will remain open or will be shut shut off SC Automatic sash device Error message relating to the automatic sash device S1 Test function The general test fun...

Страница 12: ...f controllers 24 More than 24 controllers have been connected to the communcation cable Reduce the number of controllers C3 Communication cable termination The communication cable has not been termina...

Страница 13: ...supportive flow fan Control panel faults Code Message Explanation Actions remedial measures E1 Connection lost Communication between the control panel and the fume cupboard controller has been interru...

Страница 14: ...oard 1 Fume cupboard n Supply air Extract air Communication cable The BE LCD 01 control panel is used to display and control the most critical aerodynamic and safety related functions of an EASYLAB co...

Страница 15: ...and actual values The display can be adjusted using the configuration software BE LCD 01 Operating and Installation Manual Leaflet no M375EV8 01 2015 15 5 Operation Room control Volume flow rate cont...

Страница 16: ...This enables an automatic deactivation of the increased operation after the set delay Note Whether a defaulted operating mode applies to a controller in a room depends on the configuration of the cont...

Страница 17: ...an then set any other operating mode using the control panel To resume the previous operating mode from the central BMS press the manual mode button again Manual operation can be set for a limited per...

Страница 18: ...ic signal Alarm messages Code State Explanation Actions remedial measures A1 Volume flow rate too high The volume flow rate exceeds the setpoint value Check the actuator and the controller A2 Volume f...

Страница 19: ...either no differential pressure controller on the communication cable or the room management function is not active Include a differential pressure controller Activate differential pressure control i...

Страница 20: ...ler has been interrupted For technical reasons this text always appears in English Check the cable connection the control panel and the controller E2 CP CRC failure Communication between the control p...

Страница 21: ...e for mounting the control panel onto an adapter Danger Danger of electric shock Do not touch any live components Electrical equipment carries a dangerous electrical voltage during operation Only skil...

Страница 22: ...nto place See picture on p 23 5 Align the control panel including the cable with the adapter and press it onto the fixing points until it locks into place In the process carefully push the connecting...

Страница 23: ...to the right 5 Align the control panel including the cable with the base plate and press it back onto the base plate until it locks into place In the process carefully push the connecting cable back...

Страница 24: ...alarm in case the sash is opened beyond the maximum operational sash opening Enabling the function button for Increased Operation Enabling the function button for Reduced operation Enabling the funct...

Отзывы: