Diagram control

In the following you will find information regarding the subitems
of the diagram control section (4).

Data graph

Data list

Setting the scale format

Data overview

Setting the graph format

Undo zoom

[Temperature]

[Relative humidity]

ppm

[CO2]

°C

%RH

Button

Function

Data graph

Displays the graph with the recorded
values for CO

2

, temperature and relative

humidity.

Setting the scale
format

Opens the dialog window for adjusting
the horizontal and vertical scales of the
individual channels.

Data list

Changes from the graphic display of the
measured values to a list indicating the
individual measuring points.

Data overview

Displays a summary of the gathered
data.

Setting the graph
format

Opens the dialog window for adjusting
the background colour, grid lines and
depiction of the three channels.

Undo zoom

Indicates the total view of the data
graph.

Zoom

Allows to zoom into sections of the
graph.

Cursor

Allows to select individual data points.

Adjusting the data logger settings

Click the   icon on the toolbar (2) to open the 

Data logger

settings

 dialog window.

Data logger settings

Time settings

Current time

Date format

DD/MM/YY

Sampling setting

Sampling rate

Second

Recording time

1day 4hours 5minutes 36seconds

OK

Cancel

Starting recording

Automatically

Manual

CO2 alarm settings

Limit value alarm

Alarm Value

1. Time settings

The current system time is displayed in the 

Current time

field.
The date format can be set as 

year/month/day

month/day/

year

 or 

day/month/year

.

2. Sampling setting

The 

Sampling setting

 field prompts the device to record the

measured values at a certain frequency. In the combo box
on the left, you can enter a specific value for the sampling
rate. From the combo box on the right, you can select the
time unit. The 

Recording time

 combo box shows the

maximum recording time (depends on the set sampling
rate).

3. Starting recording

The 

Manual

 and 

Automatic

 selection options allow for the

user to start data logging immediately after exiting the
setting window (

Automatic

) or to manually start at a later

point in time by pressing the respective button (

Manual

).

4. CO2 alarm settings

If the 

Limit value alarm

 checkbox is checked, the user can

enter the alarm value (in ppm). If the respective measured
value is beyond the alarm limit, the device triggers an
alarm.

Info

Prior to setting the logger to recording mode, the
logger settings [Data logger settings] must always be
made first! In this process, it is irrelevant whether the
settings are changed or remain the same. All saved
data are permanently deleted when set-up is complete.
To save those data before they are deleted, click

Cancel

 and then download the data.

Содержание 3.510.205.015

Страница 1: ...TRT BA BZ30 TC2019 23 001 EN BZ30 EN OPERATING MANUAL CO AIR QUALITY DATA LOGGER...

Страница 2: ...g the device Always store the manual in the immediate vicinity of the device or its site of use Warning Read all safety warnings and all instructions Failure to follow the warnings and instructions ma...

Страница 3: ...electrical voltage Before any work on the device remove the mains plug from the mains socket and the battery from the device Hold onto the mains plug while pulling the power cable out of the mains so...

Страница 4: ...al from 1 second to 12 hours USB interface for measuring data transfer PC analysis software Device depiction 1 3 4 6 7 2 8 12 11 9 14 10 15 13 16 17 5 DC 6 V 1A No Designation 1 Main display 2 USB int...

Страница 5: ...te If you store or transport the device improperly the device may be damaged Note the information regarding transport and storage of the device Transport The supplied Li ion batteries are subjects to...

Страница 6: ...mode and to save the entered value press the ESC button once you have set the correct date and time Info You can modify the order of day month and year within the supplied software For more detailed i...

Страница 7: ...t a new recording you first have to save or delete measured values already saved on the device For more detailed information please refer to the Using the software section Using the software There is...

Страница 8: ...smData format can be opened Opens the dialog window for printing files Opens the dialog window for adjusting the settings Starts data import from the selected device Opens the help Main menu In the fo...

Страница 9: ...rs 5minutes 36seconds OK Cancel Starting recording Automatically Manual CO2 alarm settings Limit value alarm Alarm Value 1 Time settings The current system time is displayed in the Current time field...

Страница 10: ...ime of the logger data file the endpoint of the graph time scale is used as the end time for the logger data file 2 Setting the vertical scale Scaling of the vertical axis is executed in the Vertical...

Страница 11: ...dow Use the left and bottom scrollbars to shift the pane 2 Zooming out Click the Undo zoom button to show the total view Displaying individual data points There are two options to display individual d...

Страница 12: ...not be disposed of with the household waste at the end of its life You will find collection points for free return of waste electrical and electronic equipment in your vicinity The addresses can be ob...

Страница 13: ...Trotec GmbH Grebbener Str 7 D 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com...

Отзывы: