Important Safety Instructions
SAVE THESE INSTRUCTIONS
This manual contains instructions and warnings that should be followed during the installation,
operation and storage of all Tripp Lite UPS Systems. Failure to heed these warnings may affect
your warranty.
Statement of Intended Use
Tripp Lite Medical-Grade UPS Systems are intended to support and protect non-medical and
medical computer equipment and medical devices that require leakage current reduction, surge
protection, voltage regulation, line noise filtering and battery backup during power outages and
generator testing, both inside and outside patient care areas. Tripp Lite’s Medical-Grade UPS
Systems come with hospital-grade plugs and receptacles that reduce leakage to below 100μA.
Product Information
AGOM350KSR57HG
AGSM700PSR3HG
AGSM1200PSR3HG
AC Input Voltage
120
120
120
AC Input Voltage Range
83–147
81–143
81–143
Input Amp(s)
2.3
5.5
9.2
Output Amps(s)
1.9
5.8
6.25
Rated Power (VA/Watts)
330VA/225W
700VA/450W
1000VA/750W
Frequency
50/60 Hz
50/60 Hz
50/60 Hz
No. of Phases
Single
Single
Single
Class Type
Class I
Class I
Class I
Plug/Connector Type
NEMA 5-15P
NEMA 5-15P
NEMA 5-15P
Operation
Continuous
Continuous
Continuous
AC Input Protection (Qty./type/rating)
Input breaker
(2 x 3.5A), thermal
Input breaker
(2 x 10A), thermal
Input breaker
(2 x 15A), thermal
Max Leakage Current (uA)
Less than 100uA
Less than 100uA
Less than 100uA
Weight
25.0 lbs
44.2 lbs
46.4 lbs
Known Contraindication
None
None
None
UPS Location Warnings
Relevant Symbols
Denotes that information in the manual should be reviewed before use.
Used to signal a warning that a statement is particularly important and could pose a safety
risk.
Warning-Dangerous Voltage
No sitting
Denotes general warning sign
No stepping on surface
Refer to Instruction Manual/Booklet
No pushing