TX7410 Operation Manual
50
Time
Min. value
Max. value
T1 Trigger pulse width
T2 Measurement start and setup
T3 Trigger waiting time
Measurement time——fast
——middle
——slow
Calculating and hanlding time
Display time
Measurement delay
10us
200us
0us
16ms+9T
80ms+9T
320ms+4T
10ms
15ms
0ms
---
2ms
Next valid trigger will be memorized
since this measurement.
20T at f<1kHz
20T at f<1kHz
320ms+9T at f
≥
1kHz
---
---
60s
Figure 6-3 Time diagram
6.3 Electrical Characteristics
6.3.1 DC Isolated Outputs
Each DC output (pins 1 through 11, pins 19 through 21, pins 30 through 31) is isolated
using an open collector output opto-coupler. The output voltage of each line is set by a
pull-up resistor on the handler interface board. The pull-up resistors can be connected to
/
TRIG
/
IDX
/EOM
Result
Time
previous valid data
valid data
hold or clea
r
T1
T2
T3
one measurement period
next period
delay
setup
measurem
ent time
calculating &
comparing
display
time
measurem
ent delay
triggger