TX7410 Operation Manual
21
open correction for all frequency points.
Note
:
Press
【
▲
】
【
▼
】
to enter short correction menu.
4. Press ESC key to cancel correction and exit to test status.
5. Press ENTER key to start correction.
4.2.9 Short correction
Press SHIFT + SHORT to enter short correction menu.
TX7410’s open correction can eliminate the effect of test cable, test fixture and stray
impedance in series with the device under test, such as lead resistance or lead inductance.
Perform the following steps to perform short correction:
1. Press SHIFT key at test status, and “SHIFT” at left corner is on.
2. Press SHORT key to select correction function, as Figure 4-4 shows on LCD.
Figure 4-4 Short spot correction
3. Select to perform spot, sweep or DCR correction through
【
▲
】
or
【
▼
】
key, as shown
in “SPOT” means only short correction for the current frequency point; while“SWEEP”
means short correction for all frequency points.
Note
:
Press
【
▲
】
【
▼
】
to enter open correction menu.
4. Press ESC key to cancel correction and exit to test status.
5. Press ENTER key to start correction.
4.2.10 Display mode
Press SHIFT + DISP to change display mode.
The instrument has three display modes:
Direct display (DIR)
--- Actual measured value is displayed;
Absolute deviation display (⊿)
– Absolute deviation between measured value and
nominal value is displayed;
Percentage deviation display (% )
– Percentage deviation between measured value
and nominal value is displayed.
Display mode only works for primary parameter, not secondary parameter.
Shift of display mode is related to tolerance mode of primary parameters:
When tolerance mode is absolute value (ABS)
→
Display mode shifts between DIR and
⊿
When tolerance mode is percentage (PER)
→
Display mode shifts between DIR and %.