TX7410 Operation Manual
2
TX7410 provides 3 display modes and 2 kinds of sorting modes:
Direct:
The actual measurement results are displayed.
Absolute deviation (ΔABS):
The difference between the measured value of the
DUT and a previously stored reference value are displayed.
Percentage deviation (Δ% ):
The difference between the measured value of the
DUT and a previously stored reference value are displayed as a percentage of the
reference value.
Component sorting:
Two comparison modes, absolute deviation and percent
deviation can be used.
4-bin sorting:
3 PASS bins, 1 FAIL bin, and 1
AUXiliary BIN.
TX7410 provides several communication interfaces which make it easy to output the
measurement results to other equipments (for example computers), or build an automatic
test system.
RS232C interface:
RS232C interface makes it easy to communicate with other
quipments. You can set the measurement functions and input the parameters through
RS232C interface instead of keyboard on front panel.
IEEE488 interface (optional):
IEEE488 interface helps instrument to build an
automatic test system with computer and other measurement instruments.
Command of RS232C and IEEE488 interfaces are programmed in the format of
common SCPI.
HANDLER interface:
By using the handler interface, the instrument can easily be
combined with a component handler, and a system controller to fully automate
component testing, sorting, and quality control data processing.
TX7410 has the file function to store the set parameters, limit parameters and list sweep
parameters as a file in the internal nonvolatile memory, so that the setups can be easily
reloaded, and the defult file will be automatically reloaded when TX7410 is turned on.
1.2 Operating Conditions
1.2.1 Power requirements
Vltage: 220V (1±10%)
Frequency: 50Hz/60Hz (1±5%)
Power: <50VA
1.2.2 Opearing temperature and humidity
Temperature: 0
°
C
~
40
°
C
Humidity: <90%RH