www.trilithic.com
Page V-4
120 DSP Operation Manual
Basic Signal Level Meter
120
Digital Channel Measurement
Deep Interleave
Compatibility
Yes
Downstream MER
40 ±2 dB @ +6 dBmV RF Input Level
34 ±2 dB @ -6 dBmV RF Input Level
Downstream BER
Method:
True BER, derived from code words not from MER
Standard:
ITU J.83 annex A, B, C
Range:
1 E-7 to 1 E-9 @ -6 dBmV RF Input Level
Symbol Rates
≥
0.64 msps; ≤ 7 msps
*In-service, non-scrambled standard channels only
Minimum Optimal
Input Level for
Full Range
+10 dBmV
(although not optimal, signals below this level can be measured)
Dynamic Range
50 dB
Resolution
< 0.5 dB
Carrier-to-Noise Measurement
Max Number of
Carriers
10
High/Low Delta
Resolution
0.1 dB
Scan
Video, audio, pilot, and digital carriers
Tilt Measurement
Содержание 120 DSP
Страница 17: ...www trilithic com 120 DSP Basic Signal Level Meter Section I The Basics ...
Страница 57: ...www trilithic com 120 DSP Basic Signal Level Meter Section II Setup Menu ...
Страница 128: ...www trilithic com Page II 72 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 150: ...www trilithic com Page II 94 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 151: ...www trilithic com 120 DSP Basic Signal Level Meter Section III Autotest Menu ...
Страница 152: ...www trilithic com Page III 2 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 154: ...www trilithic com Page III 4 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 168: ...www trilithic com Page III 18 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 169: ...www trilithic com 120 DSP Basic Signal Level Meter Section IV Troubleshoot Menu ...
Страница 170: ...www trilithic com Page IV 2 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 172: ...www trilithic com Page IV 4 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 212: ...www trilithic com Page IV 44 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 230: ...www trilithic com Page IV 62 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 279: ...www trilithic com 120 DSP Basic Signal Level Meter Section V Appendix ...
Страница 280: ...www trilithic com Page V 2 120 DSP Operation Manual THIS PAGE LEFT INTENTIONALLY BLANK Basic Signal Level Meter 1 20 ...
Страница 288: ...9710 Park Davis Drive Indianapolis IN 46235 317 895 3600 www trilithic com Made in U S A ...