enough material to cause a difference in light blockage of the sensors, or to block the sensors.
Data is used for trending wear as in section 8.2 Quantifying DR Data.
9.
Maintaining the DR-7
The DR-7 requires little maintenance except keeping the instrument clean and oil free and keeping the
LED's sensors calibrated.
9.1 – Sensor Cleaning
Eventually over time dirt and oil will build up in the sensor area. Routine cleaning will eliminate
possibilities of sensor reading errors caused by dirty or oily sensors. Dirt deposits can lead to obstruction
of light generated by the LED's, which in turn will increase DL and DS readings.
To clean the sensor area, follow the procedure below:
1.
Close the DR-7 program and shutdown the Windows operating system.
2.
Once the operating system is shut down, turn off the DR-7 and unplug the power cord.
3.
Remove the sensor cable coming out the top of the optics holder assembly by loosening the set screw
with a 0.05-inch Allen wrench. The set screw is located on the left side of the optic holder assembly.
4.
Gently pull upwards on the optics cable, removing the optics.
NOTE:
Ensure the position of the optics are correctly inserted the same way when re-installing into
the block, otherwise DL and DS readings will be inverted. A letter L and S are stamped into the nylon
insert. The L should be positioned into the left side of the optics holder socket.
5.
Clean the two optical sensors with heptane or isopropyl alcohol and dry them by gently blowing them
off with air. If the optical sensors come out of their sockets, they can be reinserted with the short lead
placed into the upper socket. The socket is designed to only allow the sensor to be inserted one way.
6.
Ensure the Optics Assembly is clean by lifting the gate and cleaning the area where the precipitator
tube is inserted, paying attention to thoroughly cleaning the light source area.
7.
Reinsert the optics making sure that they are fully seated in the optics block. Ensure that the L
marking on nylon insert is facing to left-hand side of the instrument.
8.
Tighten the small Allen set screw on the side of the block to hold the optics in place.
9.
Turn the Instrument back ON and follow the directions given on the screen or in section 7.3.1 –
OPTICAL CALIBRATION
of the manual to conduct optical light source calibration.
NOTE: Never remove the optics block from the magnet assembly or tamper with the adjustment
screws located on the top and sides of the optics block. Doing so will misalign the optics with the
light source. The optics are aligned and calibrated during the unit production assembly process. If this
happens contact Trico for support.