Manifold Systems
Start Up & Checking Procedures
Start up & checking procedures
The manifold is pre-programmed (per page 27) and tested
before it is shipped. You may, however, wish to modify
some of the programming (see page 16). The unit has
been designed to allow some programming to be simply
and safely altered in the field.
1.
Start with all portable bulk vessels and cylinders
turned off and with zero pressure supplied to the
manifold cabinet. To conduct the initial start-up
testing of the manifold, it is simpler and faster if
the circuit board is switched from the standard
‘cycling view’ (this is the mode in which it is
shipped) to the ‘global view’. To do this, the
cover must be removed from the circuit board and
the top #1 dip switch must be switched from the
RUN mode to the PROGRAM mode.
2.
Turn on the 120 VAC to the unit. For model
CCU the display should illuminate showing all
zeros for the Line Pressure (LIN), Left Bank
Pressure (LEFT) and Right Bank Pressure
(RIGHT). The intermediate (INT) and
emergency reserve (EMER) displays should also
display zeros. The INT and EMER displays will
always display zeros on the global view screen of
CC models and will not display when the mode is
switched to the cycling view. (CC models). The
INT and EMER displays are only activated on LL
models. Both the left & right bank Red
(Depleted) LED’s should be illuminated. Both
the left and right bank Green (In Use) and Yellow
LED’s (Ready) should be extinguished.
3.
SLOWLY open one cylinder valve on the left
bank. The left bank pressure gauge (inside the
cabinet) and the text display (on the outside of the
cabinet) should show the full pressure of the
cylinder. The Red (Depleted) LED for the left
bank should have extinguished leaving only the
Green (In Use) LED illuminated.
4.
SLOWLY open one cylinder valve on the right
bank. The right bank pressure gauge (inside the
cabinet) and the digital display (on the outside of
the cabinet) should show the full pressure of the
cylinder. The Red (Depleted) LED for the right
bank should have been extinguished and the
Yellow (Ready) LED should have illuminated.