Alarm & Information Codes
Error Codes, Alarm Codes & Information Codes (continued)
Code
Displayed
Message Displayed
Explanation
Alarm 01
Line pressure low
Message is displayed and low line pressure relay activated whenever the
line pressure is below the programmed low line pressure alarm set point.
Alarm 02
Line pressure high
Message is displayed and high line relay activated whenever the line
pressure is above the programmed high line pressure alarm set point (high
line pressure alarm is triggered).
Alarm 03
Secondary supply in use
Message is displayed and secondary in use relay activated when the
manifold has switched over to the secondary bank. Clears when tank is
replaced.
Alarm 04
Emergency reserve in use
Message is displayed and emergency in use relay activated whenever the
intermediate pressure is below the programmed emergency in use
pressure alarm set point. (LL models only).
Alarm 05
Emergency reserve pressure low
Message is displayed and emergency low relay activated whenever the
emergency pressure is below the programmed emergency reserve low
pressure alarm set point. (LL models only).
Info 01
Economizer in use – see manual
Message is displayed whenever the ready bank pressure exceeds the
service bank pressure by 50 psig or more. The logic compares the service
and ready bank pressures once a second. (LL models only).
XX% Remains
(CC models only - except for N2O & CO2 services). It calculates the
percent of gas remaining in the service bank.
Definitions and clarifications
Alarm Code
– Alarm conditions per NFPA 99C and Z7396.1 guidelines.
Error Code
– Messages that provide diagnostic information to assist in resolving system problems.
Information Codes
– Messages that provide information regarding the operation of the system.