17
Tri-Tech
Medical Inc.
Gene
sys
®
Manifold Systems
General Maintenance & Error Codes
General Maintenance
Control Cabinet
Headers & Pigtail
Daily
Record line and bank pressures
Observe nitrous oxide and carbon dioxide systems
for cylinder frosting or surface condensation. Should
excessive condensation or frosting occur it may be
necessary to increase manifold capacity.
Monthly
Check regulators, compression fittings
and valves for external leakage.
Check valves for closure ability.
Alternate line regulator in use (if dual).
Inspect valves for proper closure.
Check pigtails for cleanliness, flexibility, wear,
leakage, kinked, pinched or twisted and thread
damage. Replace damaged pigtails immediately.
Inspect header check valve outlets for closure ability.
Annually
Check relief valve pressures
Check regulator seats.
Every 4 years
Replace all pigtails
Error Codes, Alarm Codes & Information Codes
Code
Displayed
Message Displayed
Explanation
Error 01
Left bank sensor out of range
This condition is activated when the left sensor’s raw readings are at
either extreme. Can be caused by a disconnected, wired incorrectly, bad,
or over-pressurized sensor.
Error 02
Right bank sensor out of range
This condition is activated when the right sensor’s raw readings are at
either extreme. Can be caused by a disconnected, wired incorrectly, bad,
or over-pressurized sensor.
Error 03
Intermediate pressure out of range
This condition is activated when the intermediate sensor’s raw readings
are at either extreme. Can be caused by a disconnected, wired incorrectly,
bad, or over-pressurized sensor. (LL models only).
Error 04
Emergency reserve out of range
This condition is activated when the emergency sensor’s raw readings are
at either extreme. Can be caused by a disconnected, wired incorrectly,
bad, or over-pressurized sensor. (LL models only)
Error 05
Left bank pressure high
In CC models this message is displayed whenever the left inlet bank
pressure exceeds 3,000 psi. In LL models this message is displayed
whenever the left inlet bank pressure exceeds 500 psi.
Error 06
Right bank pressure high
In CC models this message would be displayed if the right inlet bank
pressure exceeds 3,000 psi. In LL models this message would be
displayed if either inlet bank pressure exceeds 500 psi.
Error 07
Intermediate pressure high
Message is displayed whenever the intermediate pressure exceeds 500 psi.
(LL models only).
Error 08
Emergency reserve pressure high
Message is displayed whenever the high pressure emergency reserve bank
pressure exceeds 3,000 psi. (LL models only).
Error 09
Line sensor noise detected
Message is displayed if a gas board detects noise in the signal from it’s
digital sensor. Noise is detected if the protocol checksums do not match.
Error 10
Line sensor failed to respond
Message is displayed if a line sensor is not responding.
Error 11
Line sensor is disconnected
Message is displayed whenever a line sensor is disconnected.
Error 12
Secondary supply leak detected
Message is displayed when a leak is detected in the secondary bank. (CC
models only).
Error 13
Emergency reserve leak detected
Message is displayed when a leak is detected in the emergency reserve
bank. (LL models only).
Error 14
Gas type mismatch
This error code is triggered by a mismatch in gas type between the line
sensor and user selected gas type in setup of the manifold circuit board.