EMC GUIDANCE
EMC GUIDANCE
EMC Guidance
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments
.
Warning
:
Don’t be near the active HF surgical equipment and the RF shielded room of an ME
system for magnetic resonance imaging, where the intensity of EM disturbances is high.
Warning
:
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
Warning: Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
Warning
:
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the
equipment TMB-2085, including cables specified by the manufacturer. Otherwise, degradation of
the performance of this equipment could result.
Technical description
:
1
.
All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE
with regard to electromagnetic disturbances for the excepted service life.
2
.
Guidance and manufacturer’s declaration - electromagnetic emissions and Immunity
Table 1
Guidance and manufacturer’s declaration - electromagnetic emissions
RF emissions
CISPR 11
Group 1
Class B
Not applicable
Compliance
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
RF emissions
CISPR 11
Emissions test
Not applicable