TROUBLESHOOTING
Battery powered mode: 2x AAA batteries 3V
About 13.5-21.5 cm or 13.5-23.5 cm
Approx.102 g (Excluding the batteries)
2x AAA batteries, user manual
Pressure: 5°C-40°C within ±3 mmHg (0.4 kPa)
Pulse value: ±5%
Rated cuff pressure:
0 mmHg ~ 299 mmHg (0 kPa ~ 39.9 kPa)
Measurement pressure:
SYS: 60 mmHg ~ 230 mmHg (8.0 kPa ~ 30.7 kPa)
DIA: 40 mmHg ~ 130 mmHg (5.3 kPa ~ 17.3 kPa)
Pulse value: (40-199)beat/minute
Power supply
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Working condition
Storage & transportation
condition
Measurement perimeter
of the wrist
Weight
External dimensions
Attachment
A temperature range of : +5°C to +40°C
A relative humidity range of 15% to 90%, non-condensing,
but not requiring a water vapour partial pressure greater
than 50 hPa
An atmospheric pressure range of: 700 hPa to 1060 hPa
Temperature: -20°C to +60°C
$UHODWLYHKXPLGLW\UDQJHRIQRQFRQGHQVLQJ
at a water vapour pressure up to 50hPa
Digital LCD V.A.43 mm x 46 mm
Approx.85.7 mm × 60.8 mm × 24.7 mm (Excluding the cuff)
SPECIFICATIONS
NOTE: If the product still does not work, contact Transtek Customer Service.
Under no circumstance should you disassemble or attempt to repair the unit
by yourself.
PROBLEM SYMPTOM
CHECK THIS
REMEDY
Error
message
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.If the problem
persists,contact the retailer or our customer
service department for further assistance.
Refer to the warranty for contact information
and return instructions.
Warning
message
“out ” shows
Out of measurement
range
Relax for a moment. Refasten the cuff and
then measure again. If the problem persists,
contact your physician.