background image

6VDC 4×AAA batteries

Approx.230g(Excluding the batteries and cuff)

A01

Digital LCD V.A.72.8mm × 74.2mm

Approx.134.8mm×100.2mm×64.8mm

4×AAA batteries,user manual,AC adapter  

5V      1A

Battery powered mode: 

AC adaptor powered mode:

(Please only use the recommended  AC 
adaptor model). 

    This section includes a list of error messages and frequently 

asked questions for problems you may encounter with your blood  
pressure monitor. If the products not operating as you think it 
should, check here before arranging for servicing.

PROBLEM SYMPTOM

CHECK THIS

REMEDY

No power

Low
batteries

Error
message

Display will not 

light up.

Batteries are exhausted. Replace with new batteries

Insert the batteries 

correctly

Replace with new batteries

Batteries are inserted
incorrectly.

Display is dim or 

show

Batteries are low.

E 1 shows

E 3 shows

E 4 shows

The treatment of the
measurement failed.

AC adaptor is inserted
incorrectly.

Insert the AC adaptor 

tightly

E 2 shows

The cuff is too tight 
or too loose.

Refasten the cuff and then
measure again.

The monitor 
detected motion or talking 
during measuring.

Movement can affect the
measurement.Relax for a 
moment and then 
measure again.

Relax for a moment and
then measure again.

The measurement 
process does not 
detect the pulse 
signal.

Loosen the clothing on
the arm and then 
measure again. 

Warning
message

Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.

“OUT ” shows

Out of measurement
range

TROUBLE SHOOTING

SPECIFICATIONS

About 22 cm to 32 cm, 22cm to 42cm,
22cm~45cm

Type BF applied part

WARNING: No modification of this equipment is allowed.

Power supply

Display mode

Measurement mode

Oscillographic testing mode

Measurement range

Measurement perimeter

of the upper arm

Weight

External dimensions

Attachment

Mode of operation

Continuous operation

Degree of protection

Protection against 

ingress of water

Accuracy

Normal working condition

Storage & transportation

condition

Software Version

Pressure:
5

-40

within±3mmHg(0.4kPa) 

Pulse value:±5%

Rated cuff pressure: 
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure: 
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa) 
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute

IP21 It means the device could protected against 
solid foreign objects of 12.5mm and greater, and 
protect against vertically falling water drops.

Device Classification

Battery Powered Mode: 
Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment

A temperature range of :+5°C to +40°C 
A relative humidity range of 15% to 90%, 
non-condensing, but not requiring a water vapour 
partial pressure greater than 50 hPa
An atmospheric pressure range of : 
700 hPa to 1060 hPa

Temperature:-20°C to +60°C

A relative humidity range of ≤ 93%, non-condensing, 

at a water vapour pressure up to 50hPa

28

29

Содержание TMB-2079

Страница 1: ...079 Please do read the user manual carefully and thoroughtly so as to ensure the safe usage of this product and keep the manual well for further reference in case you have problems version 1 0 Guangdo...

Страница 2: ...ENT 20 Recall the Records Delete the Records INFORMATION FOR USER 24 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 26 What are systolic pressure and diastolic pressure What is the standard bl...

Страница 3: ...e user manual labeling or other component They are the requirement of standard and using Symbol for THE OPERATION GUIDE MUST BE READ Symbol for MANUFACTURER Symbol for SERIAL NUMBER Symbol for TYPE BF...

Страница 4: ...t in prolonged impairment of patient blood circulation When measurement please avoid compression or restriction of the connection tubing CAUTION The device cannot be used with HF surgical equipment at...

Страница 5: ...ce in accordance with the information provided in the ACCOMPANYING DOCUMENTS Wireless communications equipment such as wireless home network devices mobile phones cordless telephones and their base st...

Страница 6: ...T 5V 1A Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batteries in fire Batteries may explode or leak Remove batteries if the device is...

Страница 7: ...p can be assigned to each record that is stored in the memory The setting range of the year 2020 2099 time format 24H 1 When the monitor is off Long press MEM button it will display the time Then pres...

Страница 8: ...USER button will decrease time 3 After the trends start time and end time are set the LCD will disply do nE and then it will turn off Select the User ID Before you start the measurement please select...

Страница 9: ...our test information Search your test information BEFORE YOU START BEFORE YOU START Install the App and Pair Up Download the MedM Health app from APP Store or Google Play Install the APP and register...

Страница 10: ...time on the same arm or as directed by a physician 6 Take the measurement in a silent room The cuff should maintain at the same level as the right atrium of the heart Please sit comfortably Do not cro...

Страница 11: ...data for the desired user Take User 1 for example Install the APP and register an account Then set your personal information Gender Birthday Height Weight Name and so on N 5 Press the START STOP butto...

Страница 12: ...ecall the Records 3 Then show the latest measurement records you can press the MEM or SET button to get the record you want The current No is No 2 The corresponding time is A M 06 00 The corresponding...

Страница 13: ...h dE LE YES the display will show dE LE YES User ID 3 Press START STOP button to confrim deleting the LCD will display do nE and then turn off If you did not get the correct measurement you can delete...

Страница 14: ...he reusable cuff with water and never immerse the cuff in water Tips for Measurement Within 1 hour after dinner or drinking Within 20 minutes after taking a bath In a very cold environment Immediate m...

Страница 15: ...andard blood pressure classification Irregular Heartbeat Detector CAUTION The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heart beat was detected during...

Страница 16: ...en measure again Warning message Relax for a moment Refasten the cuff and then measure again If the problem persists contact your physician OUT shows Out of measurement range TROUBLE SHOOTING SPECIFIC...

Страница 17: ...d found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residen...

Страница 18: ...edical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labelling and information to be su...

Страница 19: ...differential mode 0 5 kV 1 kV 2 kV common mode Conduced RF IEC61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 V 0 15 MHz 80 MHz 6 V in ISM...

Отзывы: