6VDC 4×AAA batteries
Approx.230g(Excluding the batteries and cuff)
A01
Digital LCD V.A.72.8mm × 74.2mm
Approx.134.8mm×100.2mm×64.8mm
4×AAA batteries,user manual,AC adapter
5V 1A
Battery powered mode:
AC adaptor powered mode:
(Please only use the recommended AC
adaptor model).
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the products not operating as you think it
should, check here before arranging for servicing.
PROBLEM SYMPTOM
CHECK THIS
REMEDY
No power
Low
batteries
Error
message
Display will not
light up.
Batteries are exhausted. Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Display is dim or
show
Batteries are low.
E 1 shows
E 3 shows
E 4 shows
The treatment of the
measurement failed.
AC adaptor is inserted
incorrectly.
Insert the AC adaptor
tightly
E 2 shows
The cuff is too tight
or too loose.
Refasten the cuff and then
measure again.
The monitor
detected motion or talking
during measuring.
Movement can affect the
measurement.Relax for a
moment and then
measure again.
Relax for a moment and
then measure again.
The measurement
process does not
detect the pulse
signal.
Loosen the clothing on
the arm and then
measure again.
Warning
message
Relax for a moment.
Refasten the cuff and then
measure again. If the
problem persists, contact
your physician.
“OUT ” shows
Out of measurement
range
TROUBLE SHOOTING
SPECIFICATIONS
About 22 cm to 32 cm, 22cm to 42cm,
22cm~45cm
Type BF applied part
WARNING: No modification of this equipment is allowed.
Power supply
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Measurement perimeter
of the upper arm
Weight
External dimensions
Attachment
Mode of operation
Continuous operation
Degree of protection
Protection against
ingress of water
Accuracy
Normal working condition
Storage & transportation
condition
Software Version
Pressure:
5
℃
-40
℃
within±3mmHg(0.4kPa)
Pulse value:±5%
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
IP21 It means the device could protected against
solid foreign objects of 12.5mm and greater, and
protect against vertically falling water drops.
Device Classification
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour
partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
28
29