EMC Guidance
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments
Warning
:
Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic
resonance imaging, where the intensity of EM disturbances is high.
Warning
:
Use of this equipment adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally.
Warning: Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.”
Warning
:
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment
TMB-1776-B,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
Technical description
:
1
,
all necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to
electromagnetic disturbances for the excepted service life.
2
,
Guidance and manufacturer’s declaration -electromagnetic emissions and Immunity
Table 1
Guidance and manufacturer’s declaration - electromagnetic emissions
RF emissions
CISPR 11
Group 1
Class [ B ]
Class A
Comply
Compliance
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
RF emissions
CISPR 11
Emissions test
Guidance and manufacturer’s declaration – electromagnetic Immunity
Immunity Test
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV for power supply lines
±0.5 kV, ±1 kV differential mode
0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.0 % UT;
1 cycle and 70 % UT; 25/30 cycles;
Single phase: at 0°.0 % UT; 250/300 cycle
30 A/m
50Hz/60Hz
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Table 2
Compliance level
Electrostatic
discharge
(ESD)
IEC 61000-4-2
30 A/m
50Hz/60Hz
Power frequency
magnetic field
IEC 61000-4-8
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
±1 kV signal input/output
100 kHz repetition frequency
IEC 60601-1-2
Test level
0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles; Single phase: at 0°.
0 % UT; 250/300 cycle
±2 kV, ±4kV, ±8 kV,
±15 kV air
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
±0.5 kV, ±1 kV,±2 kV common mode
±0.5 kV, ±1 kV differential mode
±0.5 kV, ±1 kV,±2 kV common mode
Conduced RF
IEC61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 V
0,15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Radiated RF
IEC61000-4-3
28
29
EMC GUIDANCE
EMC GUIDANCE