EMC GUIDANCE

EMC GUIDANCE

28

29

EMC Guidance

1)This product needs special precautions regarding EMC and needs to be 

installed and put into service according to the EMC information provided,

and this unit can be affected by portable and mobile RF communications 

equipment.

2)* Do not use a mobile phone or other devices that emit electromagnetic 

fields, near the unit. This may result in incorrect operation of the unit.

3)Caution: This unit has been thoroughly tested and inspected to assure 

proper performance and operation!

4)* Caution: This machine should not be used adjacent to or stacked with 

other equipment and that if adjacent or stacked use is necessary, this 

machine should be observed to verify normal operation in the configuration in 

which it will be used.

Table 1 

Guidance and manufacturer’s declaration – electromagnetic emissions

RF emissions 
CISPR 11

Group 1 

Class B 

Class A

Complies

Compliance

Harmonic emissions 
IEC 61000-3-2

Voltage fluctuations/ 
flicker emissions  
IEC 61000-3-3  

RF emissions 
CISPR 11

Emissions test

Electromagnetic environment - guidance

The device is intended for use in the electromagnetic environment specified below. The customer or the 

user of the device should assure that it is used in such an environment.

The device uses RF energy only for its internal 

function. Therefore, its RF emissions are very low 

and are not likely to cause any interference in nearby 

electronic equipment.

The device is suitable for use in all establishments,

other than domestic and those directly connected to 

the public low-voltage power supply network that 

supplies buildings used for domestic purposes.

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

Immunity test

 

±8 kV contact
±15 kV air 

±8 kV contact
±15 kV air 

 

±2 kV 

power supply lines: 

line(s) to line(s): ±1 kV  

line(s) to earth: ±2 kV 

0%U

T;

 0.5 cycle

At 0°, 45°, 90°, 135°,

180°,225°,270° and 315°

0%U

T

 ;  1 cycle 

and 

70%U

T

 ; 25/30 cycles

Single phase: at 0°

0% U

T

 ; 300 cycle

30 A/m

50Hz/60Hz

NOTE      U

is the a.c. mains voltage prior to application of the test level. 

Table 2 

Compliance level

Electrostatic 
discharge (ESD) 
IEC 61000-4-2 

Electromagnetic 

environment - guidance

30 A/m

50Hz/60Hz

Power frequency 

(50Hz/60Hz) 

magnetic field 

IEC 61000-4-8 

Voltage dips, 

short interruptions

and voltage

variations on 

power supply 

input lines

IEC 61000-4-11 

Electrical fast 
transient/burst 
IEC 61000-4-4 

Surge
IEC61000-4-5

input/output lines:

±1 kV

The device is intended for use in the electromagnetic environment specified below. The customer or the 

user of the device should assure that it is used in such an environment.

IEC 60601 test level

Floors should be wood, concrete 

or ceramic tile. If floors are 

covered with synthetic material, 

the relative humidity should be at 

least 30%.

±2 kV 

power supply lines: 

100 kHz repetition

frequency

line(s) to line(s): ±1 kV  

100 kHz repetition

frequency

Mains power quality should be 

that of a typical commercial or 

hospital environment.

Mains power quality should be 

that of a typical commercial or 

hospital environment.

Mains power quality should be 

that of a typical commercial or 

hospital environment.

Power frequency magnetic fields 

should be at levels characteristic 

of a typical location in a typical 

commercial or hospital 

environment.

0% U

T

 ; 0.5 cycle

At 0°, 45°, 90°, 135°,

180°,225°,270° and 

315°

0% U

T

 ;  1 cycle 

and 

70% U

T

 ; 25/30 cycles

Single phase: at 0°

0% U

T

 ;300 cycle

Содержание TMB-1598-BS

Страница 1: ...age of this product and keep the manual well for your further reference in case you have problems Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB 1598 BS Wrist Type Version 1 0 G...

Страница 2: ...blood pressure at home compared to the hospital Is the result the same if measuring on the right wrist TROUBLESHOOTING 24 SPECIFICATIONS 25 CONTACT INFORMATION 26 Safety Information Start Measurement...

Страница 3: ...devices patients with pre eclampsia premature ventricular beats atrial fibrillation peripheral arterial disease and patients undergoing intravascular therapy or arterio venous shunt or people who rec...

Страница 4: ...and remove it from your wrist If the cuff pressure reaches 40 kPa 300 mmHg the unit will automatically deflate Should the cuff not deflate when pressures reaches 40 kPa 300 mmHg detach the cuff from t...

Страница 5: ...he average value of blood pressure User 1 Indicate the blood pressure level Blood pressure level indicator Start measurement and save the results for user 1 User 2 Start measurement and save the resul...

Страница 6: ...idelines The typical service life of the new and unused batteries is 54 measurements for the operation time is 60s 1 2 3 It is important to set the clock before using your blood pressure monitor so th...

Страница 7: ...steps 2 and 3 to confirm Positioning symbol on or off 6 Repeat steps 2 and 3 to confirm the measurement unit 7 After confirming the meausrement unit the LCD will display donE then display all the set...

Страница 8: ...st cuff around your wrist leaving no extra room between the cuff and your skin If the cuff is too loose the measurement will not be accurate Sit comfortably with your tested wrist resting on a flat su...

Страница 9: ...l display User ID donE and then turn off Then you can start your measurement now 2 1 Adjust the zero LCD display Inflating and measuring Display and save the result The year date and time will dispaly...

Страница 10: ...cinity of equipment marked with the following symbol And TMB 1598 BS may interfering vicinity electrical equipment Sensitive people including pregnant women pre eclamptic and those who implanted medic...

Страница 11: ...confirm Then press button to check the records of the selected user 3 1 If you did not get the correct measurement you can delete all the results for the selected user by following steps Take user 1 f...

Страница 12: ...t Immediate measurement after tea coffee smoking When talking or moving your fingers When you want to discharge urine Maintenance In order to get the best performance please follow the instructions be...

Страница 13: ...ths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down It is ok for both wrists but there will be some different results for different people We su...

Страница 14: ...your monitor with your mobile device Please check below items Bluetooth is ON App Collector is ON Both devices are within the transmission distance of Bluetooth Power supply Battery powered mode 2 AA...

Страница 15: ...anagement to medical devices EN 1041 2008 Information supplied by the manufacturer of medical devices EN 60601 1 2006 IEC 60601 1 2005 A1 2012 Medical electrical equipment Part 1 General requirements...

Страница 16: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Страница 17: ...quency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths f...

Страница 18: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Отзывы: