This section includes a list of error messages and frequently 

asked questions for problems you may encounter with your blood 

pressure monitor. If the products not operating as you think it 

should, check here before arranging for servicing.

24

25

SPECIFICATIONS

TROUBLESHOOTING

PROBLEM SYMPTOM CHECK THIS

REMEDY

No power

Low

batteries

Error

message

Batteries are exhausted. Replace with new batteries

Insert the batteries 
correctly

Replace with new batteries

Batteries are inserted

incorrectly.

Display is dim or

display 

Batteries are low.

E 01 shows

The cuff is too tight 

or too loose.

Refasten the cuff and then

measure again.

E 02 shows

E 03 shows

E 04 shows

Relax for a moment and

then measure again.

EE XX,shows on 

the display.

A calibration error

occurred.

(XX can

be some digital 

symbol,such as 01,

02,etc., if this similar

situation appear, all 

belong to calibration 

error.)

Retake the measurement.

If the problem persists,

contact the retailer or our

customer service 

department for further 

assistance.Refer to the 

warranty for contact 

information and return 

instructions.

The measurement 

process does not detect 

the pulse signal.

Loosen the clothing on 

the wrist  and then  

measure again.

The treatment of the 

measurement failed.

Display will not 
light up.

The monitor detected
motion,talking or the

pulse is too poor 

while measuring.

Relax for a moment

and then measure again.

Follow the instructions in 

user manual and measure

again. 

“oUt ” shows

“E 12 ” shows

Out of measurement

range.

Failure of pairing up your

monitor with your mobile

device.

Please check below items:

-Bluetooth is ON.

-App Collector is ON.

-Both devices are within 

the transmission distance 

of Bluetooth.

Power supply

Battery powered mode: 

2*AAA alkaline batteries (3V DC)

Display mode

Digital  LCD  V.A.45mm×33mm

Measurement mode

Oscillographic testing mode

Measurement range

Accuracy

Normal working condition

Storage & transportation

condition

Measurement perimeter

of the wrist

About 13.5cm-21.5cm

Weight

Approx.100g(Excluding the dry cells)

External dimensions

Attachment

Approx.86mmx66mmx22mm
2×AAA batteries,user manual  

Mode of operation

Continuous operation

Degree of protection

Type BF applied part

Protection against 

ingress of water

Software version

Device classification

A01
Internally Powered ME Equipment

WARNING: No modification of this equipment is allowed.

Pressure:

5°C-40°C within±3mmHg(0.4kPa) 

Pulse value:±5%

Rated cuff pressure: 

0mmHg~299mmHg(0kPa ~ 39.9kPa)

Measurement pressure: 

SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa) 

DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)

Pulse value: (40-199)beat/minute

IP22: The first number 2: Protected against 

solid foreign objects of 12,5mm Ф and 

greater. The second number: Protected 

against vertically falling water drops when 

enclosure titled up to 15º. Vertically falling 

drops shall have no harmful effects when the 

enclosure is titled at any angle up to 15º on 

either side of the vertical.

Temperature:-20°C to +60°C

A relative humidity range of ≤ 93%, 

non-condensing, at a water vapour pressure 

up to 50hPa

A temperature range of :+5°C to +40°C 

A relative humidity range of 15% to 90%, 

non-condensing, but not requiring a water 

vapour partial pressure greater than 50 hPa

An atmospheric pressure range of : 

700 hPa to 1060 hPa

Содержание TMB-1598-BS

Страница 1: ...age of this product and keep the manual well for your further reference in case you have problems Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB 1598 BS Wrist Type Version 1 0 G...

Страница 2: ...blood pressure at home compared to the hospital Is the result the same if measuring on the right wrist TROUBLESHOOTING 24 SPECIFICATIONS 25 CONTACT INFORMATION 26 Safety Information Start Measurement...

Страница 3: ...devices patients with pre eclampsia premature ventricular beats atrial fibrillation peripheral arterial disease and patients undergoing intravascular therapy or arterio venous shunt or people who rec...

Страница 4: ...and remove it from your wrist If the cuff pressure reaches 40 kPa 300 mmHg the unit will automatically deflate Should the cuff not deflate when pressures reaches 40 kPa 300 mmHg detach the cuff from t...

Страница 5: ...he average value of blood pressure User 1 Indicate the blood pressure level Blood pressure level indicator Start measurement and save the results for user 1 User 2 Start measurement and save the resul...

Страница 6: ...idelines The typical service life of the new and unused batteries is 54 measurements for the operation time is 60s 1 2 3 It is important to set the clock before using your blood pressure monitor so th...

Страница 7: ...steps 2 and 3 to confirm Positioning symbol on or off 6 Repeat steps 2 and 3 to confirm the measurement unit 7 After confirming the meausrement unit the LCD will display donE then display all the set...

Страница 8: ...st cuff around your wrist leaving no extra room between the cuff and your skin If the cuff is too loose the measurement will not be accurate Sit comfortably with your tested wrist resting on a flat su...

Страница 9: ...l display User ID donE and then turn off Then you can start your measurement now 2 1 Adjust the zero LCD display Inflating and measuring Display and save the result The year date and time will dispaly...

Страница 10: ...cinity of equipment marked with the following symbol And TMB 1598 BS may interfering vicinity electrical equipment Sensitive people including pregnant women pre eclamptic and those who implanted medic...

Страница 11: ...confirm Then press button to check the records of the selected user 3 1 If you did not get the correct measurement you can delete all the results for the selected user by following steps Take user 1 f...

Страница 12: ...t Immediate measurement after tea coffee smoking When talking or moving your fingers When you want to discharge urine Maintenance In order to get the best performance please follow the instructions be...

Страница 13: ...ths before beginning will be better for measuring Advice Relax yourself for 4 5 minutes until you calm down It is ok for both wrists but there will be some different results for different people We su...

Страница 14: ...your monitor with your mobile device Please check below items Bluetooth is ON App Collector is ON Both devices are within the transmission distance of Bluetooth Power supply Battery powered mode 2 AA...

Страница 15: ...anagement to medical devices EN 1041 2008 Information supplied by the manufacturer of medical devices EN 60601 1 2006 IEC 60601 1 2005 A1 2012 Medical electrical equipment Part 1 General requirements...

Страница 16: ...hat supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines l...

Страница 17: ...quency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths f...

Страница 18: ...960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3...

Отзывы: