30
31
TROUBLESHOOTING
SPECIFICATIONS
PROBLEM SYMPTOM CHECK THIS
REMEDY
Error
message
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.If the problem
persists,contact the retailer or our customer
service department for further assistance.
Refer to the warranty for contact information
and return instructions.
Warning
message
out shows
Out of measurement
range
Relax for a moment. Refasten the cuff and
thenmeasure again. If the problem persists,
contact your physician.
Battery powered mode: 2*AAA batteries
About 13.5cm-21.5cm
Approx.102g(Excluding the batteries)
2*AAA batteries,user manual
Pressure:5°C-40°C within±3mmHg(0.4kPa)
Pulse value:±5%
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Power supply
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Working condition
Storage & transportation
condition
Measurement perimeter
of the upper arm
Weight
External dimensions
Attachment
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%, non-condensing,
but not requiring a water vapour partial pressure greater
than 50 hPa
An atmospheric pressure range of : 700 hPa to 1060 hPa
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
Digital LCD display V.A.32mmx45mm
Approx.85mm×67mm×23mm(Excluding the cuff)