![Transducer Techniques LCA-9PC Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/transducer-techniques/lca-9pc/lca-9pc_operators-manual_1151163005.webp)
-
5-
Set E4 jumper
to select 5V excitation output. Leave out for 10V excitation output.
Excitation Voltage
5V
10V
E4
jumper
install
none*
* Factory default setting.
Set twin E1 jumpers
to select normal or reversed input polarity for a positive voltage
output.
Signal polarity
Normal
Reversed
E1
jumper
a + a*
b + b
* Factory default setting.
Set the E3 jumper position f
to select a 4-20 mA output. Do not set that jumper for
other output ranges.
Signal output
0 to ±5V 0 to ±10V 0-16 mA 0-20 mA
4-20 mA
E3
jumper
no f*
f
* Factory default setting
Set E2 gain jumpers
for combinations of excitation voltage (5V or 10V), load cell sensi-
tivity (0.5 to 10.0 mV/V), and signal output (0-5V, 0-10V, 0-16 mA, 0-20 mA or 4-20 mA).
Exc.
Output
Load cell sensitivity in mV/V
V
o
I
o
0.5
1.0
1.5
2.0
2.5
3.0
4.0
10.0
5V
1-5V
4-20 mA
f
a, e
b, d b, c, e* c, e
c
b, e
e
0- ±5V 0-20 mA
f
a, e
b, d b, c, e* c, e
c
b, e
e
0- ±10V
NA
a, d, f
f
a, c, e a, e b, d, e b, d b, c, e
b
10V
1-5V
4-20 mA
a, e b, c, e
c
b, e
b
b
e
none
0- ±5V 0-20 mA
a, e b, c, e
c
b, e
b
b
e
none
0- ±10V
NA
f
a, e
b, d b, c, e* c, e
c
b, e
e
* Factory default setting
Set E3 and E8 jumpers
to set low-pass filter action, which applies to both voltage and
current outputs. At the frequencies shown (0.1 to 30 kHz), signal and noise amplitudes
will be down by -6 dB (or a factor of 2).
-6 dB point in kHz
0.1
0.3
1.0
3.0
10
30
E3
jumper
a
b
c*
d
e
none
E8
jumper
a
b
c*
d
e
none
* Factory default setting