F19
Key-in Zero
Allows you to key-in known zero
calibration value in case of
memory loss in the field.
Press
ZERO
key to begin
sequence
F20
Key-in Span
Allows you to key-in a known
span calibration value in case of
memory loss in the field.
Press
ZERO
key to begin
sequence
F21
Factory
Reset
This sub-menu will reset all
parameters to the default
settings.
USE WITH CAUTION.
Press the
ZERO
key
twice to
execute.
F23
DA Output
Type
Selects the Analog Output type:
4-20mA or 0-10V
Aout-C
√
Aout-V
F24
DA Output
Enable
Allow/Forbid analog output
(4-20mA or 0-10V)
DA-dIS
DA-EnA
√
F25
DA Digital
Adjust
Adjusts the analog output value.
Press
ZERO
key to begin
sequence
-7-
3.2 User menu
Changing communication mode and formula is in this menu.
A menu
Name/Code Description Code/Value
A1
Baud Rate
Selects the baud rate for data
transmission through the serial
port.
1200 2400
4800 9600
19200
√
A2
Data Bits and
Parity
Selects the number of data bits
and parity of serial transmission.
8n
√
7O
7E
7n
A3
Mode of
Serial
Transmission
Selects when data will be sent
out of the serial port to a printer
or computer:
“C”=Continuous mode
“d”= Demand mode
d
√
C
A4
Display
Check
Actuates the function that
illuminates all digit segments,
decimal points, and LCD
annunciators in a test sequence.
Press
ZERO
key to begin
sequence
A5
ID No. Entry
Settings the Signal resource,
when more than one device in
one net.
[01-32]
01
√
A20
Enable
ZERO key in
unstable
Allow/Forbid use ZERO key to
clear zero when the system is in
unstable state.
rZ-dIS
√
rZ-EnA
-8-