F4
Zero Range
Selects the range within which
the scale may be zeroed. Note
that the indicator must be in
standstill to zero the scale.
1.9
2
√
100
F5
Motion Band
Sets the level at which motion is
detected by comparing the
present display update with the
previous one.
1d
√
3d
5d 10d
F6
Digital Filter
AUTO: Auto Digital Filter
FAST: Fast Filter
8: More Stability
16: Most Stability
AUTO FAST
8
16
√
F6A
Digital Filter
Average Value Filter
16
32
√
48
64
F7
Overload
Limit
Selects the desired formula which
determines the point at which the
indicator shows overload. All
selections are based on the
primary unit selected in F8.
“FS”=Full scale in primary units
Overload display “
□□□□□□
”
FS
FS+2%
√
FS+1d
FS+9d
F8
Calib. Unit
Selects the primary base unit to
be used in the calibration
process. Also the default unit for
normal operation.
1-lb
2-kg
√
F9
Display
Divisions
Determines the desired weigh
increments. Value should be
consistent with legal
requirements.
div-1
√
div-2
div-5
-5-
F10
Decimal Pt.
Determines location of the
decimal point.
0
√
0.0
0.00 0.000
0.0000 00
F14
Disable the
lb/kg key
Allows the lb/kg key to be
disabled so that an operator
cannot accidentally press the key
and change the displayed units.
Cr-DIS
√
Cr-EnA
F15
Zero Range
when Power
on
Selects the zero range when the
system power on.
OFF
√
(
±
)10%
100%
F16
Zero
Calibration
Places the system into the zero
calibration routine.
Press
ZERO
key to begin
sequence
F17
Span
Calibration
Places the system into the span
calibration routine.
Press
ZERO
key to begin
sequence
F18
View
Calibration
Actuates the function that allows
you to view both the zero and
span calibration value. The
values displayed in this function
are valid only after Calibration
(F16 & F17) has been
successfully completed.
Press
ZERO
key to show
zero value;
Press ”
→
” key
to show span
value.
-6-