Page 6-2
4. After setting the exact value, press the NET/GROSS key to save the value.
5. If the calibration was successful, the display will show "
EndC1
" momentarily, then revert back
up to F17. At this time it is suggested that the calibration values be recorded for future use
(see Section 6.4).
6. If the calibration was
not
successful, one of the error messages below will appear. Take the
indicated action to correct the problem, then perform a new calibration.
"
Err0
" - The calibration test weight or the adjusted keyed-in weight is larger than the full
capacity of the scale. Change the calibration test weight or check the input data.
"
Err1
" - The calibration test weight or the adjusted keyed-in weight is smaller than 1% of the
full capacity of the scale. Change the calibration test weight or check the input data.
"
Err2
" - The internal resolution of the scale is not high enough to accept the calibration value.
Select a larger parameter for the Span Gain (F2). SEE APPENDIX C FOR MORE
INFORMATION.
6.4
VIEW CALIBRATION VALUES (F18)
Note:
The values displayed in this procedure are valid only after a successful calibration has been
performed using F16 and F17.
1. While in the Setup mode, scroll to "
F 18
", then scroll down once using the ZERO key to enter
View calibration menu.
2. The display will momentarily show "
CAL 0
" followed by a value. This value is the
zero
calibration value
and should be recorded in the table below. Press any key to continue.
3. The display will momentarily show "
CAL 1
" followed by another value. This value is the
span
calibration value
and should also be recorded in the table below. Press any key to return to
upper level (F18).
INDICATOR
ZERO CALIBRATION VALUE
SPAN CALIBRATION VALUE
S/N:
Table 6-1: Calibration Value Table
6.5
KEY-IN ZERO CALIBRATION VALUE (F19)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss.
A valid zero calibration value, obtained from a successful F16 calibration procedure, must
be used.
1.
While in the Setup mode, scroll to "
F 19
", then scroll down once using the ZERO key.
2.
The display will momentarily show "
CAL 0
", followed by a flashing zero. Use the four
directional keys (shown in Figure 6-1) to adjust the displayed value to the zero calibration
value.
3.
After setting the exact value, press the NET/GROSS key to save the value.
4.
The display will show "
E CAL 0
" momentarily, then revert back up to F19.