Page 4-2
NAME/CODE
DESCRIPTION
CODE/VALUE
F9
Display
Divisions
Determines the desired weight increments. Value should be consis-
tent with legal requirements.
1
√√
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0
√√
0.0
0.00 0.000
0.0000 00
F16
Zero Calibra-
tion
Places indicator into the zero calibration routine. Scrolling down with
the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F17
Span Calibra-
tion
Places indicator into the span calibration routine. Scrolling down
with the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F18
View Calibra-
tion
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F19
Key-in Zero
Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the
ZERO
key one level begins
the procedure.
Press
ZERO
key
to begin sequence
F20
Key-in Span
Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Scrolling down with the
ZERO
key one level
begins the procedure.
Press
ZERO
key
to begin sequence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menu to
the default settings. USE WITH CAUTION!
Press the
ZERO
key twice to exe-
cute.