PARAMETER
(Soft Key Text)
DESCRIPTION AVAILABLE
CHOICES
F1
Graduations
Specifies number of full-scale graduations. Value should be
consistent with legal requirements and environmental limits on the
useful system resolution. Note – default for 30 lb or 15 kg is
15000 divisions.
1000, 2000, 5000
10000,
12000
√
15000, 24000
30000, 50000
F3
Zero Track
Selects the range within which the scale will automatically zero.
Note that the scale must be in standstill to automatically zero.
Selections are in Display Divisions.
0d
0.5d
1d
2d
√
3d
5d
F5
Motion Band
Sets the level at which motion is detected by comparing the
present display update with the previous one. If motion is not
detected for two seconds or more, scale is in standstill and can
process a Print or Zero command. Maximum value varies
depending on local regulations.
1d
√
3d
5d
10d
F6
Digital Filter
Averages weight readings to produce higher stability. The higher
the filter setting, the greater the stability but the slower the
indicator’s response time. Choose 8 unless a very fast response
is needed.
8,12,16,24,
32
√
48, 64
F8
lb or kg
Sets the starting unit during power up
1 = lb
√
2 = kg
F9
Display
Divisions
Determines the desired weight increments. Value should be
consistent with legal requirements.
1
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0 0.0
0.00 0.000
0.0000
F16
Zero Calibration
Sets zero calibration. See chapter 6 for calibration instructions
F17
Span Calibration
Performs scale calibration. See chapter 6 for calibration
instructions
F21
Restore factory
default
Restores all parameters to factory defaults
F23
Scale Capacity
Selects capacity of scale
12 lb, 30 lb, 60 lb,
120 lb, 6 kg, 15
kg, 30 kg, 60 kg
17