30
www.trace2o.com
Metalyser HM5000 Benchtop
Instruction Manual v1.0
Calibration
The data calculated above is a calibration curve for the sample
being analysed. If desired, this calibration can now be applied to
new samples. To do this, some extra data is first required.
The calibration function requires the peak position (in V) to be set. Set the
max peak shift (the maximum distance from the peak position where the
peak is recognised), the min peak width for which the peak is recognised,
and the min peak height for which the peak is recognised.
Set the calibration temperature (the temperature at which the calibration
scans were carried out), and the blank offset (the peak height of the blank
scan. This may be due to unavoidable contaminants in the buffer, for
example natural levels of Pb in NaCl) and the units of the calibration.
Click ‘Calculate’ and the HM5000 will use the data to create a calibration curve.
Thereafter, if ‘Use Calibration’ is checked, the HM5000 will use the calibration data
that is stored and calculate the concentration of the sample in the units specified.