Page 20
8.3 Calibration Messages
1. If a temperature calibration has been successfully performed, the
WP-80M
will display the following message and the offset value of the
sensor. For example:
Calibrate OK
Offset= 1.0
°°°°
c
2. If a temperature calibration has failed, the
WP-80M
will display the
following message, and the failed offset value of the sensor. For
example…
Calibrate Failed
Offset= 10.5
°°°°
c
8.4 Manual Temperature Setting
If the temperature sensor is not connected, the temperature of the sample
solution must be set manually for accurate pH measurements. A separate
thermometer will be required for this.
1. Switch the meter on.
2. Measure the temperature of the sample.
3. Select Temperature Calibration (
→
→
→
→
F1:Cal.
→
→
→
→
F2:Temp
).
4. The current temperature setting is now displayed. For example…
> 25.0< Man Temp
↑
↑
↑
↑↓
↓
↓
↓
:Set F1:Save
5. Press the
and
keys until the display shows the temperature of
the sample.
6. Press the
key to save the temperature value.
Alternatively, press the
key to quit and retain the current setting.