Page 17
6.
mV/ORP Calibration
The mV/ORP section is factory calibrated. There is no user-calibration
facility for this mode.
7.
Relative mV/ORP Calibration
Select Relative mV/ORP mode when measurements relative to a known
standard are required. Calibration of the Relative mV/ORP mode is simply
a matter of zeroing the reading when the sensor is in the known standard.
1. Plug the Redox sensor into the
pH/mV
socket. Temperature
compensation is not applied in Relative mV/ORP mode, so the
temperature sensor does not need to be connected.
2. Switch the meter on.
3. Select Relative mV/ORP Mode (
→
→
→
→
F3:Mode
→
→
→
→
F2:Rel mV
).
The display should now be showing Relative mV/ORP on the top line
with absolute mV/ORP or date and time on the bottom line. For
example…
1000mVR 25.0
°°°°
c
660mV F4:Zeros
↔
1000mVR 25.0
°°°°
c
31/12 12:00:00
Press
to alternate between absolute mV/ORP or the clock on the
bottom line.
4. Remove the wetting cap from the Redox sensor.
5. Rinse the sensor in distilled water and blot dry.
6. Place the Redox sensor into a sample of the known standard. Ensure
that the platinum tip and reference junction are both covered, as per
the diagram over the page.