13
Calibration of FPA Series Balance Internal/Linearity calibration
1) Remove all objects from pan, make sure the balance show zero,
2) Press CAL key for 5 seconds, it shows --CAL- , and then show ------, press ZERO/TARE key for 5 seconds, it shows
--CAL- again, then will show the first calibration weights, twinkling,
3) Put on the weights on the pan accordingly, it shows- , when it is stable, it shows the weights, remove the weights, it
shows- , and then shows the second calibration weights , twinkling,
4) Put on the weights on the pan accordingly, it shows- , when it is stable, it shows the weights, remove the weights, it turn
to zero automatically, the calibration is complete.
Note: We suggest you do external calibration too when you begin to use the balance. External calibration
1) Remove all objects from pan, make sure the balance show zero,
2) Press CAL key for 5 seconds, it shows --CAL- , and then show ------,and then show the calibration weights, twinkling,
3) Put on the weights , it shows ------ , when it is stable, it shows the weights, remove the weights, it turns to zero
automatically, the calibration is complete.