                                          6. Weigh the gravity hammer in the air.                     

HI

 

                                                       

e.g.: The quality of gravity hammer is 6.8135g in the air

     

6.8135 

 

Short press

CAL

                                    7.    It records the value weighed in the air                 

LO

 

                                                                            And indicating gravity hammer measured in liquid   

6.8135 

g

 

                                          8

Take off the weighed sample                                 

LO

 

                                            It indicates gravity hammer measured in liquid           

0.0000 

g

 

                                          9. Weigh the gravity hammer in liquid                         

LO

 

                                                                   

e.g.: The quality of gravity hammer is 1.3518g in liquid

         

1.3518 

g

 

Short press

CAL

                                    10. It records the weigh value in liquid                           

OK               

                                                      In the meantime calculate the weighed liquid density     

00.99707 

g

d

 

  If you want to measure different liquid’s density , please repeat step 5-10.   

  In the liquid density measure mode, if you want to reset volume of gravity hammer, please press

MODE

to re-enter the gravity hammer 

volume settings.   

Exit the density mode : Long press

MODE

and choose the normal weigh mode 

NORMALg 

 

                                                                                                                                                               

 

 

Operation Menu

 

Switch on, enter the weigh condition; press

TARE

+

ZERO

at the same time, 

in 3 seconds it displays Cod0000. Press

PRINT

to change the position, press

TARE

to scroll the numbers 

or choose the parameters , press

ZERO

to confirm and enter the next step .   

Code   

Mark display   

Description 

Setting range   

Cod0001 

Zero-x.x 

Zero range   

0.0d to 6.0d 

Stdy-x.x 

Start sensitive range 

0.0d to 6.0d 

Sens--X 

Inner sensitivity grade 

1,2,3,4,5,6 

Filt-X 

Filter radix   

1,2,3,4,5,6,7

1 weakest

7strongest 

Speed 

Weighing speed 

1

2

3

    3 fastest   

BL-xxx 

LCD back light mode 

ON

OFF

AUT 

BEPxxx 

Beeper switch 

ON, OFF 

MOdRec 

Whether record the weigh mode   

0

Always goes to normal weigh mode when switch on;   

1

record  the weigh mode  last used before switch off  ,  enter 

this mode directly when switch on .     

TAdj 

Display temperature correction   

1.9

 to 

+1.9

 

Baudxx 

Baud rate   

12

1200

24

2400

48

4800

96

9600

 

Con—xxx 

Communication method   

Non

No  communication 

CON

:  continuous

Sty

communicate when stable

Key

communication when press 

[print]

SOFT

software exchange 

Txxx

communicate by 

time

xxx is the set time, unit is second. 

dAT Con 

Output Format 

Prt:

 Our Company Standard Output Format   

Com:

 Common Data Format 

Содержание FPA Series

Страница 1: ... series Electronic Balance Operation Manual TPS SCALES VIET NAM JSC THINH PHAT SCALES JOINT STOCK COMPANY 57 D1 Street Ward 25 Binh Thanh District HCMC Vietnam TEL 0084 28 62 888 666 FAX 0084 28 3512 7966 ...

Страница 2: ...ghing Operation 3 Basic Weighing Function 3 Calibrate Adjust 4 Application Function 4 Counting Function 5 Percentage Weigh Function 5 Density Weigh Function 6 Operation Menu 8 Communication Parameters 9 Unit Conversion 10 Technical Specification 10 Repairing and Maintenance 12 ...

Страница 3: ...gh precision sensor technology magnetic balance it is the ideal device for fast accurate weighing widely used in medical research institute schools enterprise road construction and defense department etc LCD backlight display easy and clear Operator Highly smart tare and back to zero within the whole weighing scope overload display Multi mode choice counting percentage weigh density weigh With RS ...

Страница 4: ...e shaking or vibrating plate to make the balance stable Recommend to use aseismic marble Avoid to put balance in the place with big temperature change or air flow such as the place with direct sunlight or cold air outlet Please use the separate power socket to avoid the interruption by other device Please do not put anything on the scale when power on the balance Make balance adjust itself to the ...

Страница 5: ...aration Connect the power supply when the panel display press button ON OFF to switch on the balance Pre heat Time In order to ensure the precision of weighing please preheat the balance completely reference to technical parameter form page 19 20 then start the following operation Example Basic weigh operation balance already pre heated we take 210g 0 0001g balance as an example button command ste...

Страница 6: ...dicated scale Display the calibrated scale value after 5 sec 200 0000 g 4 Take off the scale complete calibration 0 0000 g Application Function Operation Choose the Mode button command steps panel display Long press MODE Enter the choice for mode Normal weigh mode NORMAL g Count mode COUNT pcs Percent weigh mode PERCENT Density measure mode DENSITY d Density measure mode is an option if you need p...

Страница 7: ...x number 20 pcs Remove the samples 0 pcs Weigh the unknown numbers in this e g 200pcs 200 pcs In Count Mode if you want to reset the radix number value please press MODE to reenter the Count Mode Exit the Count Mode Long press MODE choose the normal weigh mode NORMALg Percent Weigh Function The end user can make a certain weight as definition 100 then make other weight displayed as the percent of ...

Страница 8: ... SOLID d Short press CAL 2 Confirm the solid density measure mode Enter the liquid temperature coefficient settings 01 00000 d The user can set the liquid temperature coefficient value on his own please refer to Form1 Press PRINT to change the position press TARE to scroll down the number press CAL to confirm enter the solid density measure situation 3 The balance reminding to weigh in the air HI ...

Страница 9: ... 13 0 99941 27 0 99654 The weigh method for liquid density Use density measure device the volume of planed standard sample must be known The user need to enter the volume of standard sample manually The latest data entered will be stored for further use Step 1 To measure the liquid density please weigh the standard sample in the air Step 2 Then weigh the standard sample in the liquid want to be me...

Страница 10: ...s TARE ZERO at the same time in 3 seconds it displays Cod0000 Press PRINT to change the position press TARE to scroll the numbers or choose the parameters press ZERO to confirm and enter the next step Code Mark display Description Setting range Cod0001 Zero x x Zero range 0 0d to 6 0d Stdy x x Start sensitive range 0 0d to 6 0d Sens X Inner sensitivity grade 1 2 3 4 5 6 Filt X Filter radix 1 2 3 4...

Страница 11: ...by the operator Blank 3 2Common Data Format ST 0000 0000 g CR LF LF ST Steady 0 Value 0000 0000 g CR LF US Unsteady 0 Value Number Display Area Unit Display Area Carriage Return Line Feed 4 The list of command which the communicating device can send in software exchange mode ASCII code No Command Function 1 Sample 2 P Print 3 T Tare 4 Z Back to Zero 5 U Unit Unit Conversion Mark Unit Conversion ra...

Страница 12: ...evel Weigh scope 0 100g 0 110g 0 120g 0 200g 0 210g 0 310g value of scale division 0 0001g Standard weights value 100g 100g 100g 200g 200g 300g Repeatable tolerance standard tolerance 0 0003g Linear tolerance 0 0003g Stable time 4s Pre heat time 60 120 min Operation temperature 20 2 5 Pan diameter 90mm Overall dimensions 325 x 205 x 305mm L W H Weigh chamber dimensions 180mmX175mmX200mm L W H Net ...

Страница 13: ...3100g 4100g 5100g 6100g Precision level Weigh scope 0 1100g 0 2100g 0 3100g 0 4100g 0 5100g 0 6100g value of scale division 0 01g Standard weights value 1000g 2000g 3000g 4000g 5000g 5000g Repeatable tolerance standard tolerance 0 02g Linear tolerance 0 02g Stable time 3s Pre heat time 60 90 min Operation temperature 20 7 5 Pan diameter ф160mm Overall dimensions 325mm 205mm 100mm L W H Net weight ...

Страница 14: ...m the date it sold out Damages not caused by misuse or vandalism all belong to warranty exclude the following case Balance out of warranty Damage caused by misuses such as long time under sunshine or receive corrosion The user does not obey the operation manual such as overloading the balance Without our authorized department or technician repairing the user disassembly the balance on his own Any ...

Страница 15: ...ws the second calibration weights twinkling 4 Put on the weights on the pan accordingly it shows when it is stable it shows the weights remove the weights it turn to zero automatically the calibration is complete Note We suggest you do external calibration too when you begin to use the balance External calibration 1 Remove all objects from pan make sure the balance show zero 2 Press CAL key for 5 ...