8
6. Weigh the gravity hammer in the air.
〖
HI
〗
(
e.g.: The quality of gravity hammer is 6.8135g in the air
)
6.8135
g
Short press
[
CAL
]
7. It records the value weighed in the air
〖
LO
〗
And indicating gravity hammer measured in liquid
6.8135
g
8
、
Take off the weighed sample
〖
LO
〗
It indicates gravity hammer measured in liquid
0.0000
g
9. Weigh the gravity hammer in liquid
〖
LO
〗
(
e.g.: The quality of gravity hammer is 1.3518g in liquid
)
1.3518
g
Short press
[
CAL
]
10. It records the weigh value in liquid
OK
In the meantime calculate the weighed liquid density
00.99707
g
d
※
If you want to measure different liquid’s density , please repeat step 5-10.
※
In the liquid density measure mode, if you want to reset volume of gravity hammer, please press
[
MODE
]
,
to re-enter the gravity hammer
volume settings.
Exit the density mode : Long press
[
MODE
]
and choose the normal weigh mode
NORMALg
Operation Menu
Switch on, enter the weigh condition; press
[
TARE
]
+
[
ZERO
]
at the same time,
in 3 seconds it displays Cod0000. Press
[
]
to change the position, press
[
TARE
]
to scroll the numbers
or choose the parameters , press
[
ZERO
]
to confirm and enter the next step .
Code
Mark display
Description
Setting range
Cod0001
Zero-x.x
Zero range
0.0d to 6.0d
Stdy-x.x
Start sensitive range
0.0d to 6.0d
Sens--X
Inner sensitivity grade
1,2,3,4,5,6
Filt-X
Filter radix
1,2,3,4,5,6,7
;
1 weakest
,
7strongest
Speed
Weighing speed
1
,
2
,
3
;
3 fastest
BL-xxx
LCD back light mode
ON
,
OFF
,
AUT
BEPxxx
Beeper switch
ON, OFF
MOdRec
Whether record the weigh mode
0
:
Always goes to normal weigh mode when switch on;
1
:
record the weigh mode last used before switch off , enter
this mode directly when switch on .
TAdj
Display temperature correction
-
1.9
to
+1.9
Baudxx
Baud rate
12
:
1200
;
24
:
2400
;
48
:
4800
;
96
:
9600
Con—xxx
Communication method
Non
:
No communication
;
CON
: continuous
;
Sty
:
communicate when stable
;
Key
:
communication when press
[print]
;
SOFT
:
software exchange
;
Txxx
:
communicate by
time
,
xxx is the set time, unit is second.
dAT Con
Output Format
Prt:
Our Company Standard Output Format
Com:
Common Data Format