Section 7 – Troubleshooting

 

 

41

Section 7. –  Troubleshooting 

7.1 

General Error Messages from the easyCHEM 770  

Error Message 

Possible Causes 

Remedy 

Fact Cal Failure 
maybe Inaccurate 

The EEPROM chip which 
contains the factory 
calibration information has 
failed. 

The unit must be returned to TPS for 
service. 

EEPROM Failure 
Factory repair

 

User calibration settings have 
been lost or corrupted. 

Switch the meter OFF and switch back 
ON. If the problem persists, return the 
unit to TPS for service. 

7.2 General 

Faults 

Symptom Possible 

Causes 

Remedy 

Main Pool Pump does 
not operate when 
easyCHEM
 770  
activates output 

1.  Pool pump power lead not 

correctly connected to 

POOL PUMP

 socket. 

2.  Output Fuse is blown. 

Ensure that the Pool Pump’s power 
plug is firmly connected to the 

POOL 

PUMP

 output socket. 

Replace output fuse, as per details in 
section 7.7. If fuses continue to blow, 
your Pool Pump may be drawing too 
much power, or it may be faulty. 

Not Dosing. “PUMP” 
Error message 
displayed. 

Excess Dosing pump current 
detected. 

Press Tech Menu to reset error.  If error 
re-occurs dosing pump will require 
service. 

 

7.3 ORP 

Troubleshooting 

Symptom Possible 

Causes 

Remedy 

ORP reading inaccurate 

1.  Platinum tip of ORP 

sensor dirty. 

 
 

 
 
 

2.  Reference junction of pH 

sensor dirty. 

Clean Platinum tip of ORP sensor by 
dipping into a 1:10 mix of Hydrochloric 
acid for 10 to 20 seconds, or sand with 
800 to 1200 grade wet-and-dry 
sandpaper. 

If problem persists, adjust the set point 
to compensate for variation in reading. 

Dip pH sensor into a 1:10 mix of 
Hydrochloric acid for 10 to 20 seconds. 

Re-calibrate ORP in ZoBell solution 
(section 5.3). 

If problem persists, adjust the set point 
to compensate for variation in reading. 

ORP reading seems low 
compared to Chlorine 
test kit 

Pool water is out of balance 

Have sample of pool water tested by a 
pool professional to check its chemical 
balance. Check at least the following… 

  Cyanuric Acid (sunscreen) 

 Alkalinity 

 pH 

(your 

easyCHEM 770

) tests 

this). 

  Free and Combined Chlorine levels. 

Once the above have been rectified, it 
may be necessary to Super-chlorinate 
your pool to oxidise undesirable 
Chloramine compounds. Your pool 
professional can advise on this. See 
section 6.5 for details on using the 

easyCHEM 770

 for super-chlorinating 

your pool. 

 

Содержание easyCHEM 770

Страница 1: ...easyCHEM 770 Pool Management System Chlorine Acid controller with dosing pumps and timer for pool pump easyCHEM 770 User Manual Date 06 Sep 2007 Author AB Version 1 31...

Страница 2: ...td ABN 30 009 773 371 4 Jamberoo Street Springwood Brisbane Australia 4127 Phone 07 32 900 400 International 61 7 32 900 400 Fax 07 3808 4871 International 61 7 3808 4871 Email tps tps com au Web www...

Страница 3: ...h the Pool Pump OFF manually 19 3 7 Pumping to Waste 20 3 8 Backwashing the Pool Filter 21 3 9 easyCHEM 770 Display Messages 22 Section 4 Peristaltic Dosing Pumps 23 4 1 Lubricating the Pump Tubing 23...

Страница 4: ...on installing the easyCHEM 770 It is a duplicate of the information found on the easyCHEM 770 Installation Guide card Section 3 Operating Instructions The Operating Instructions provide details on ad...

Страница 5: ...Section 1 Introduction 5 1 1 easyCHEM 770 Display and Controls 1 2 easyCHEM 770 Connectors...

Страница 6: ...menus and data entry screens without saving changes Press to activate Auto Control mode The easyCHEM 770 will automatically add Chlorine and Acid as required Press to place the easyCHEM 770 into Stan...

Страница 7: ...Part No 121381 Probe Injection Tee 50mm Part No 117904 50 to 40mm reduction bushes x 2 Part No NPRB5040 Peristaltic Pump Input and Output Tube 4m x 2 Part No 116360 ZoBell ORP Solution Part A 200mL Pa...

Страница 8: ...e bottom of the chemical drums to the dosing pumps is 2 metres 4 The mains power lead from the pool pump reaches the easyCHEM 770 without straining 5 The mains power lead from the easyCHEM 770 reaches...

Страница 9: ...mm diameter vent hole to avoid suction build up IMPORTANT INFORMATION REGARDING ACID Always use dilute Hydrochloric Acid Add 5 Litres of Acid to 20 to 25 Litres of water NEVER ADD WATER TO ACID Using...

Страница 10: ...CHEM 770 controller on a sturdy wall or post using minimum 3mm or 1 8 stainless steel bolts screws or masonary anchors The easyCHEM 770 controller MUST be installed in a sheltered location Connect the...

Страница 11: ...unction has been included for those installers who wish to check the operation of the dosing pumps when installation is complete The pump priming function can also be used to prime the peristaltic pum...

Страница 12: ...reaches 700mV The control band stops the pump being rapidly switched on and off when the reading is close to the Set Point thereby increasing the pump s life 2 The exact ORP Set Point to achieve the c...

Страница 13: ...control band If the Set Point is 7 40pH the Acid pump will start when the rises above 7 60pH and stop again when the reading reaches 7 40pH The control band stops the pump being rapidly switched on a...

Страница 14: ...lorine and Acid as required Auto mode should be selected for normal operation 2 Press to place the easyCHEM 770 into Standby mode In this mode the unit will measure the ORP and pH levels of the pool b...

Страница 15: ...wer bill and for the environment by reducing the electricity being used 1 Press to display the Pump Timer menu Pool Pump Timed 2 Select Timed from the menu and press to select the default Timer progra...

Страница 16: ...is selected and press 4 The easyCHEM 770 will now prompt you to enter the Winter Start date Winter Start 15 May Enter Use the and keys to set the date and press Use the and keys to set the month and...

Страница 17: ...und if selected The Custom program is covered in section 3 4 4 To operate the pump 24 hours per day select Continuous This is covered in the next section 3 4 3 To select the Continuous program The Con...

Страница 18: ...is now displayed AM OFF 10 00 Set Enter Press or to set the hours for when the pool pump will turn OFF and press Press or to set the minutes for when the pool pump will turn OFF and press 5 The PM ON...

Страница 19: ...o be switched OFF by the Timer cycle 2 To stop the pump press select Timed or OFF and press 3 6 To Switch the Pool Pump OFF manually The pool pump can be switched OFF manually using the easyCHEM 770 a...

Страница 20: ...screen is displayed Set filter to Waste Enter The pool pump is now switched OFF Set the control on your filter to its Pump to Waste setting then press 3 The pool pump is now pumping pool water to was...

Страница 21: ...control on your filter to its Backwash setting then press 3 The pool pump will now backwash the filter to waste for 60 seconds A countdown is provided on the screen Backwashing 60 Filter 4 When the c...

Страница 22: ...ction 3 3 sOFF Emergency ShutOFF timer has been tripped See section 6 2 OFF Dosing Pump is off line due to 0 secs of ON time being set See section 6 3 2 OFF OVR for reading Dosing Pump is off line due...

Страница 23: ...acing the Pump Tubing Replace the Pump Tubing if it gets worn begins to leak or pumping volume is reduced The Pump Tubing may be replaced every year as a precautionary measure To remove the old Pump T...

Страница 24: ...ng to the compression fittings See the procedure detailed in section 2 4 The power to the easyCHEM 770 can now be re connected and switched on 4 3 PUMP Error Message The easyCHEM 770 monitors the dosi...

Страница 25: ...s select Calibrate then select Cal Temp The Temperature Calibration screen is now displayed For example 25 0o c Temp Cal 26 0o c Enter 6 Use the and keys to enter the same temperature as measured by t...

Страница 26: ...sfully performed the easyCHEM 770 will display the following message and the offset of the sensor For example Calibrate OK 0 1o c Offset 2 If a temperature calibration has failed the easyCHEM 770 will...

Страница 27: ...et and the ORP Temperature sensor into the ORP Probe socket if not already connected 4 Remove the wetting cap from the pH sensor Rinse both sensors in distilled water Shake off as much water as possib...

Страница 28: ...ymmetry or the Slope being outside these limits please consult the Troubleshooting guide section 7 4 for possible remedies 2 All calibration information is retained in memory when the easyCHEM 770 is...

Страница 29: ...d the ORP Temperature sensor into the ORP Probe socket if not already connected 4 Remove the wetting cap from the pH sensor Rinse both sensors in distilled water Shake off as much water as possible Bl...

Страница 30: ...lue For example Cal Failed Offset 70mV 3 If the reading is below 100 mV or above 400 mV the easyCHEM 770 will not even attempt calibration as this indicates a user error such as incorrect calibration...

Страница 31: ...the year then use the and keys to set the year Press the key to move to the hours then use the and keys to set the hours Press the key to move to the minutes then use the and keys to set the minutes 3...

Страница 32: ...within the limit ORP ShutOFF 70mins Enter Use the and keys to set the required time from 1 to 240 minutes Press and hold the or key to scroll rapidly 3 Press to save the ShutOFF timer setting for the...

Страница 33: ...one or more of the following Re fill the Chlorine or Acid drum if they were empty Check and re calibrate the ORP and or pH sensors to ensure the reading is actually correct Extend the ON time if the p...

Страница 34: ...psed The ON and OFF cycles for ORP and pH run sequentially This allows time for mixing for each of the chemicals and avoids the problem of the Chlorine affecting the pH reading or the Acid affecting t...

Страница 35: ...means that the easyCHEM 770 will add Chlorine when the reading gets too low Too High means that the easyCHEM 770 will reduce the ORP when it gets too high This is rare and not normally used in pool an...

Страница 36: ...ding gets too low Press or until the desired setting is selected and press 5 The current Control On Time is now displayed for example pH ON Time 1 min Enter Press or until the desired On Time is set a...

Страница 37: ...pped When the alarm function has been tripped the message Alrm flashes next to the ORP and or pH reading whichever has tripped the alarm and the easyCHEM 770 beeps Automatic control will continue even...

Страница 38: ...Section 6 User Settings 38 6 4 2 Alarm Set up Examples The following diagram illustrates the ORP alarm function of the easyCHEM 770 The Set Point is at 700 mV...

Страница 39: ...to enter the number of Litres of Chlorine that you wish to add to the pool and press The maximum dose that can be added is 25 Litres 5 The easyCHEM 770 now proceeds to add the Chlorine The display sho...

Страница 40: ...and press 4 The Reset menu is now displayed for example Calibration Full System Select Calibration and press to reset only the calibration information to factory values It will be necessary to re cali...

Страница 41: ...to reset error If error re occurs dosing pump will require service 7 3 ORP Troubleshooting Symptom Possible Causes Remedy ORP reading inaccurate 1 Platinum tip of ORP sensor dirty 2 Reference junctio...

Страница 42: ...ce if necessary Ensure that the bulb AND the reference junction are fully immersed as per diagram in section 5 2 1 Inaccurate readings even when calibration is successful Reference junction blocked Cl...

Страница 43: ...r outlet and unplug the power lead 2 Locate the fuse holder on the left hand underneath side of the unit Un screw it anti clockwise until it can be removed 3 Fully remove the glass fuse from the fuse...

Страница 44: ...Asymmetry Range 1 00 to 1 00 pH pH Slope Range 85 0 to 105 pH Buffers pH7 00 and pH4 00 Temperature Offset Range 10 0 O C to 10 0 O C Calendar Clock 24 Hour Calendar Clock Tests for valid day of the m...

Страница 45: ...ou feel that this equipment is in need of repair please re read the manual Sometimes instruments are received for repair in perfect working order This can occur where the sensor s simply requires clea...

Отзывы: