background image

 

4.2 Function 2: - Temperature Reading

 

The pump will stop running when in this function 

 
Remember:- 

It was the 

Function Key  

          that got you here!!!  

 
Ensure you have a ‘K’ type probe connected to one or both of the 
thermocouple sockets T1 or T2 (see 4.2) 
 
 
 
 
 
 
 
 
 
 
 

 
WARNING: -

 There is 

ONLY

 one correct way to connect the  

‘K’ type thermocouple plug into the socket (see 2.4). The 
thermocouple plug is designed with one thick (negative) and one 
thinner (positive) prong. Forcing the plug into the socket the 
wrong way round may result in damage to the instrument.

 

Failure to comply may result in damage to the instrument 
and will invalidate the warranty.

 

 

4.2.1 Screen 1

   

 

Displays Temperature reading of T1 in degrees 
Centigrade (ºC) or degrees Fahrenheit (ºF) 
Displays Temperature reading of T2 in degrees 
Centigrade (ºC) or degrees Fahrenheit (ºF)

 

 

Displays the Differential Temperature (Diff.) between 
T1 and T2 in ºC or ºF 
 

Pressing the Up Arrow Key         will toggle between 
ºC and ºF 
 
‘oPEn’ will be displayed if no ‘K’ type probe is 
connected to the thermocouple socket 
 

 
 
 
 

Содержание 717R

Страница 1: ...TPI 717R Combustion Flue Gas Analyser ...

Страница 2: ...n 2 Temperature Reading 4 3 Function 3 Pressure Testing 4 4 Function 4 Tightness Let By Test 4 5 Function 5 CO Build Up Test 4 6 Function 6 Date Time 5 Storing Saving Data 6 Holding Data on Screen 7 Reviewing Data 8 Printing Data 9 Turning Off Charging Appendix A Specifications Appendix B Calibration Service Appendix C Guarantee Appendix D Troubleshooting Guide Appendix E Index ...

Страница 3: ...below before returning your instrument to your distributor should an error occurs We can rectify over 75 of proposed faults over the phone TPI HELPLINE 01293 530196 opt 1 Your TPI 717R Combustion Analyser comes complete with the following items as standard TPI 717R Instrument c w Integral Magnet Rubber Boot Soft Carrying Case Sampling Probe c w Type K Thermocouple In Line Water Trap Bowl Filter c ...

Страница 4: ...2 Instrument Overview 2 1 Front View Rubber Boot Protects the instrument from accidental damage Display Large 3 Parameter Backlit Display Battery Indicator Showing battery life ...

Страница 5: ... Moves down through the Stored Data Addresses see 5 7 8 Used to HOLD Data on Screen see 6 Scrolls through Flue Gas Analysis Function Screens see 4 1 6 Starts Moves through the Let By Tightness Test see 4 4 Starts Stops the CO Build Up Test see 4 5 Allows you to choose a Stored Data Address see 5 7 8 Sends stored data to a separate infrared printer see 8 Allows you to view stored data on the displa...

Страница 6: ...e Pump Protection Filter see 2 4 3 Connections for Pressure Tubing see 2 4 4 Calibration and Info Label Displays calibration information Displays serial number Battery Compartment Holds rechargeable battery Rubber Boot Protects the instrument from accidental damage ...

Страница 7: ...2 3 Side Views Exhaust Port Port for connection of Exhaust Adaptor Infrared Window Window for sending stored data to IR Printer see 8 Rubber Boot Protects the instrument from accidental damage ...

Страница 8: ...see 4 1 T1 Socket Connection for any K type thermocouple probe see 4 2 Thermocouple Connection for any K type thermocouple probe see 4 2 T2 Socket Gas Inlet Port Connection for In Line Pump Protection Filter see 3 Pressure Port Connections for Pressure Tubing see 4 3 4 4 Pressure Port Connections for Pressure Tubing see 4 3 ...

Страница 9: ... second purge period briefly displaying the Software Version Instrument Serial Number before continuing with the countdown where upon the following screen will be displayed The instrument should ALWAYS be turned on in a clean air environment as the 30 second purge will attempt to set the Carbon Monoxide level to 0 ppm and the Oxygen to 20 9 If there is insufficient clean air in the sensor chamber ...

Страница 10: ...the Func Backlight Key and move onto Section 4 2 of the manual At any time you can activate the Backlight by holding down the Func Backlight Key for 2 seconds 4 1 Function 1 Flue Gas Analysis Ensure you have connected the Temperature Sampling Probe complete with In Line Water Trap Bowl Filter to the In Line Pump Protection Filter See 3 which should ALWAYS be connected to the Gas Sample Port see 2 ...

Страница 11: ...Analyser if used correctly If you are working with OIL then you MUST ensure that you also use the optional Oil Filter as shown below Failure to do so will result in erroneous readings If any of the filters become excessively dirty or blocked then the following screen will be displayed as a warning and no further readings will be able to be taken until either the dirty filters are replaced or the b...

Страница 12: ...e and one thinner positive prong Forcing the plug into the socket the wrong way round may result in damage to the instrument Failure to comply may result in damage to the instrument and will invalidate the warranty Pressing the Enter Key will take you through the following Flue Gas Analysis Screens 4 1 1 Screen 1 Displays Carbon Monoxide CO reading in parts per million ppm Displays calculated Carb...

Страница 13: ...r before the instrument will be able to be switched off see 4 1 7 Failsafe Turn Off 4 1 3 Screen 2 Displays Oxygen O2 reading in percentage Displays calculated Excess Air X Air figure in percentage Displays calculated Efficiency Eff figure in percentage Pressing the Up Arrow Key will toggle between the Fuel Options Pressing the Down Arrow Key will toggle between the Efficiency Options 4 1 4 Screen...

Страница 14: ...et Pressing the Up Arrow Key will toggle between ºC and ºF 4 1 7 CO above 15ppm Failsafe Protection Beep The 717R will not allow the user to either Turn the instrument OFF or to move to another Function whilst the CO level is above 15ppm A short beep will be heard should the user attempt either of the above with the CO above 15ppm This is a Failsafe Feature of the 717R to protect the instrument fr...

Страница 15: ...ne thinner positive prong Forcing the plug into the socket the wrong way round may result in damage to the instrument Failure to comply may result in damage to the instrument and will invalidate the warranty 4 2 1 Screen 1 Displays Temperature reading of T1 in degrees Centigrade ºC or degrees Fahrenheit ºF Displays Temperature reading of T2 in degrees Centigrade ºC or degrees Fahrenheit ºF Display...

Страница 16: ... Ensure you have Pressure Sampling Tube connected to one or both of the Pressure Ports see 2 2 or 2 4 4 3 1 Screen 1 Displays Pressure reading in either millibars mbar kiloPascals kPa or inches of Water inH2O Pressing the Up Arrow Key will scroll through mbar kPa and inH2O Pressing the Down Arrow Key will Zero the Pressure reading ...

Страница 17: ...f pressure 3 Press Enter Key to start the Test 4 The Initial Pressure will be stored at P1 and P2 will become the Live pressure 5 The 717R will countdown from 60 seconds 6 Once the countdown is finished the 717R will display P1 P2 and Differential P2 P1 and give you the option to Pass or Fail depending on whether the readings are within the allowable tolerance or not 7 Choosing Fail by pressing th...

Страница 18: ...readings are within the allowable tolerance or not 13 Choosing Fail by pressing the Down Arrow Key will stop the test and allow you to if required store save see section 5 or print see section 7 the readings so far 14 Choosing Pass by pressing the Up Arrow Key will take you to the next part of the test Tightness 15 Apply the required amount of pressure 16 Start the Tightness part of the test by pr...

Страница 19: ...ion 7 the readings test 4 5 Function 5 CO Build Up Test Remember It was the Function Key that got you here The CO room test function enables the 716 to monitor and log ambient CO levels in a room or office space at 1 minute interval up to a total of 50 minutes This data can be retrieved later via the optional infrared printer 1 Begin with the 717R in a fresh clean air environment 2 Press the Enter...

Страница 20: ...w you to change the Time Date and Year 2 Use the Print Recall Keys to choose the Flashing Numeric to change 3 Use the Up Down Arrow Keys to make the changes 4 Once the changes have been made press the Store Save Key 5 Press the Enter Key to exit and return back to the Date Time Screen ...

Страница 21: ...the Up and Down Arrow Keys Please Note Choosing the same Address Location on which previously Stored Data is being held will OVERWRITE the previous readings with the New Stored Data This will be the case regardless of the particular readings you are attempting to save i e should you have Combustion Readings stored at Address 1 and you Save Temperature Readings to Address 1 then the Previous Combus...

Страница 22: ...e Down Arrow Key until the 717R beeps approx 2 seconds 3 The readings will freeze the word HELD will be displayed on the bottom line to indicate you are viewing HELD readings 4 You can scroll through the other combustion screens by pressing the Enter Key repeatedly 5 The HELD readings can also be printed whilst in HOLD Mode Please follow instructions from section 7 1 6 The HELD readings can also b...

Страница 23: ...f you have chosen to view Flue Temperature or Pressure Readings then Addr will be displayed and a location number will be flashing 4 Select the required address location that you wish to review by pressing the Up and Down Arrow Keys 5 Press the Enter Key once and the Time Date of the Saved Data from the selected address location will be displayed on the screen If there is no data present at that l...

Страница 24: ... screen below left will be displayed and the Real Time rEAL readings or Last Time Calibrated Date CAL will be sent to the printer However if you have chosen to print the Stored Data Stor then FLuE will be flashing Select the required type of data Flue Temperature Pressure Tightness or CO Build Up Test that you wish to print by pressing the Up and Down Arrow Keys 3 Press the Enter Key to select the...

Страница 25: ...nt has an auto shut off after 10mins should no keys have been pressed for this period and as mentioned above that the CO is below 15ppm Should the CO be above 15ppm then the 10 minute auto shut off countdown will not begin till the CO has gone below 15ppm 9 1 Charging Plug the Charger supplied into the charger socket on the instrument see 2 4 If the instrument is turned on then a charging symbol w...

Страница 26: ... Weight 500g Casing Rubber Boot as Standard Switch Off Failsafe Exhaust Safety Spigot Conforms to EN50379 BS7967 Flue Temperature Probe Construction Pistol Grip with Stainless Steel Shaft Hose Length 2500mm Insertion Length 200mm K Type Thermocouple Accuracy 0 3 1 C Maximum Temperature 800 C Gases Range Resolution Accuracy Oxygen 0 25 0 1 0 3 Carbon Monoxide 0 10 000 ppm 1 ppm 20 ppm 3 ppm 100 ppm...

Страница 27: ...tional Part No A740 Various K Type Probes Optional See TPI Brochure Oil Filter Complete Optional Part No A773 Smoke Test Pump Optional Part No A788 Appendix C GUARANTEE Your TPI 717R Combustion Analyser is guaranteed free from defects in materials and workmanship for 6 Years subject to annual service being carried out by TPI Europe or one of its approved service centres Covered by TPI Repair parts...

Страница 28: ...ort Negative pressure readings on display Either there is suction or pressure was not zeroed in atmospheric air Re zero in atmospheric air Negative temperature readings on display Thermocouple plug has been plugged in the wrong way round Remove thermocouple plug and plug in the correct way round Flo Err is displayed Blockage in sampling probe or kink blockage in sample tube Check and rectify as ne...

Страница 29: ...ing 5 Dimensions Appendix A Display 2 1 Appendix A Efficiency 4 1 1 4 1 3 4 1 4 4 1 5 Excess Air 4 1 3 Exhaust Port 2 3 Filters Appendix B Front View 2 1 Fuel Selection 4 1 1 4 1 3 4 1 4 4 1 5 Fuels 4 1 1 4 1 3 4 1 4 4 1 5 Appendix A Func Backlight Key 4 7 84 Guarantee Appendix C Infrared Printer 1 8 Appendix B Infrared Window 2 3 8 Inlet Ports 2 2 2 4 3 1 4 1 4 2 4 3 In Line Filter 1 3 1 4 1 Appe...

Страница 30: ... 2 Recall Key 2 1 7 Rubber Boot 1 2 1 2 2 2 3 2 4 Appendix A Appendix B Scroll Enter Key 2 1 4 1 4 4 4 5 5 6 7 8 Service Appendix B Side Views 2 3 Specifications Appendix A Stabilisation 4 4 Store Save Key 2 1 Temperature Selections 4 2 Thermocouple 2 4 4 1 4 2 Appendix A Appendix B Tightness Test 4 4 Top View 2 4 Troubleshooting Appendix D Warranty Appendix B Weight Appendix A TPI Helpline 01293 ...

Страница 31: ......

Страница 32: ...fo tpicanada com Test Products International Europe Ltd Longley House International Drive Southgate Crawley W Sussex RH10 6AQ England Tel 44 0 1293 530196 Fax 44 0 1293 531870 contactus tpieurope com www tpieurope com This symbol indicates that at the end of its life this product must not be disposed of in household waste This equipment should be environmentally disposed of Please contact your sup...

Отзывы: