9. Start the stabilisation part of the test by pressing
the
Enter Key
10. The Initial Pressure will be stored at P1 and P2
will become the “Live” pressure.
11. The 717R will countdown from 60 seconds
12. Once the countdown is finished the 717R will
display P1, P2 and Differential (P2-P1) and give
you the option to “Pass” or “Fail” depending on
whether the readings are within the allowable
tolerance or not.
13.
Choosing “Fail” by pressing the
Down
Arrow Key
will stop the test and allow you to, if
required, store/save (see section 5) or print
(see section 7) the readings so far.
14.
Choosing “Pass” by pressing the
Up
Arrow Key
will take you to the next part of the
test
– Tightness
15. Apply the required amount of pressure
16. Start the Tightness part of the test by pressing
the
Enter Key
17. The Initial Pressure will be stored at P1 and P2
will become the “Live” pressure.
18. The 717R will countdown from 120 seconds