4.1.5 Screen 4
Displays CO air free calculated reading
Pressing the Up Arrow Key will toggle between
the Fuel Options
Pressing the Down Arrow Key will toggle
between the Efficiency Options
4.1.6 Screen 5
Displays Temperature reading of T1 in degrees
Centigrade (ºC)
Displays Temperature reading of T2 in degrees
Centigrade (ºC)
Displays the Differential Temperature (Diff.) between T1
and T2 in ºC
‘oPEn’ will be displayed if no ‘K’ type probe is
connected to the thermocouple socket
Pressing the Up Arrow Key will toggle between
ºC and ºF
4.1.7 CO above 15ppm (Failsafe) Protection Beep
The 717R will not allow the user to either Turn the instrument OFF or to move to
another Function whilst the CO level is above 15ppm. A short beep will be heard
should the user attempt either of the above with the CO above 15ppm.
This is a Failsafe Feature of the 717R to protect the instrument from giving false
readings when next turned on. The 717R should be run in a clean air environment
with just the In-Line Pump Protection Filter attached until the CO level drops
below 15ppm. The 717R will then be able to be moved to a different function
screen or be turned OFF.