6. MAINTENANCE
1. Be sure there is no resin or material sticking to the surface of the light guide.
2. If materials are sticking on the light guide immediately clean it with soft cloth. Keep light guide clean for better
light intensity and curing performance.
3. The light guide should be sterilized after each use. It can be safely autoclaved at 121 °C. For no more then 15
minutes. Let cool before handling.
4. The other accessories such as curing light hadpiece and eye protector should be scrubbed with pure water and
sheathed in accordance with standard disinfection procedures.
5. Keep curing light handpiece in charging station while not in use.
6. Dispose Lithium battery according to the federal and state laws.
7. The User should wear colored protection glasses to avoid eye injury due to LED light.
7. TROUBLESHOOTING
Fault
Possible Cause
Solution
The low battery indicator on the
curing light is solid red
Battery Low
Charge Battery
The overheat indicator light is
orange
The curing light has overheated
Allow to cool for 10 min
When the light is placed on the
charging base the LED emmits a
low light while not in use
Battery is very low
Disconnect battery and charge bat-
tery while disconnected from light
for 10 min or till fully charged
My light guide is very loose when
placed in the curing light
O-ring damaged or missing
Check o-ring or repalce
The light changes modes on its own Function button is depressed in
Verify the function button is not
stuck. Press and release the func-
tion button
None of the mode or status lights
are illuminating
Drained battery
Contacts are dirty on the battery
Depleted battery
Charge battery
Clean contacts
Replace battery
Charging base does not beep when
batteries are placed on the charger
No power to base.
Connect power supply to base.
The light intensity indocator on the
base illuminates red when checking
the light intensity
Light guide had debris on it.
Not enough battery power to cure
Clean light guide
Charge battery
When i try to attach the battery its
very tight and will not fully attach
Battery may have been dropped
and the threads are dented in
Repalce battery
I see a halo pattern when the light
is being used
Light guide has cracks in the fiber
Replace light guide