Page 5
ITM-MAGMETER-T500-ED2
Nominal diameter
DN 10 to DN 1200
Nominal pressure
PN 10 to PN 16
Tubing connection
Flanges DIN 2633
Measuring tube internal coating
Rubber, PTFE
Sensing electrodes
Stainless steel AISI 316L, Hasteloy
2 x measuring / 1 x indication of empty tube /1 x ground
Flow range
0,1 to 10 m/s
Measurement error
0,5% of measured value in range from 0,5 to 10 m/s
1% of measured value in range 0,1 to 0,5 m/s
Ambient temperature
-20 to 60°C
Voltage
230V AC (optionally 15-24V AC/12-34V DC)
Input power
10 VA max.
Outputs
1 x programmable multifunctional (pulse, status), isolated
(relay, load 125 V ~ /1A or 30V/2A)
1 x programmable multifunctional (pulse, status), isolated
(transistor NPN, load 30 V/50 mA max.)
1 x programmable active current (0-20 mA/ 4-20 mA,
load ≤ 500Ω)
1 x RS232
Communication
RS232, keyboard, permanently backlit graphical display,
SMS via internal GSM module
Datalogger
Capable of recording up to 100 samples
Protection
IP67: sensor / IP65: converter (compact version)
IP67/68: sensor / IP65: converter (remote version)
DN
L (mm)
Weight
(Kg)
15
150
2
20
150
2
25
150
2,5
32
150
3,5
40
150
5
50
200
6
65
200
9
80
200
11
100
250
13,5
125
250
20
150
300
25
200
350
35
250
400
41
300
500
55
350
500
65
400
600
110
500
600
120
Содержание T-500