Page 13
ITM-MAGMETER-T500-ED2
B1. Pulse setting
B2. Status report setting
Volumes 0.1 l, 1 l, 10 l, 100 l, 1 m3, 10 m3, 100 m3 and widths 10 ms, 20 ms, 40 ms, 80 ms, 160 ms, 320 ms
can be assigned to transmitted pulses.
WARNING!
The fl ow meter transmits number of pulses corresponding to liquid quantity fl owed during measuring
cycle of 640 ms. Pulse can be assigned only such volume and width so that number of transmitted
pulses during time unit is real! Incorrect setting is indicated by error message. Display shows the code
E5.
Use cursor keys to select option PULSE OUTPUTS from main menu and confi rm your selection by EXE
1.
key. A submenu will be displayed.
Select pulse width under option ms/imp1(2) and assign required volume to the pulse under option
2.
l/imp1(2).
Select option OUT1 (output to terminals 1 - 2) or OUT2 (output to terminals 3 - 4) and confi rm your selec-
3.
tion by EXE key.
Select V+ from menu (output transmits pulses only for fl ow in positive direction), V- (output transmits
4.
pulses only for fl ow in negative direction) or V (output transmits pulses for both directions) and confi rm by
EXE key.
Select option RETURN to fi nish setting.
5.
The instrument will ask if you want to save new parameters SAVE SETUP?. Select YES if they are OK. The
fl ow meter will save new parameters and switch to normal display mode. If NO is selected, parameters will not
be saved.
The instrument will switch to display mode and works with new parameters until it is switched off. When the
instrument is switched on again, it works with original parameters.
Use cursor keys to select option PULSE OUTPUTS from main menu and confi rm your selection by EXE
1.
key. A submenu will be displayed.
Select option OUT1 (output to terminals 1 - 2) or OUT2 (output to terminals 3 - 4) and confi rm your selec-
2.
tion by EXE key.
Select option COMP1(2) from menu (to assign status function to output) and confi rm by EXE key.If status
3.
function COMP1(2) is assigned to outputs, further setting of this function will be performed under option
COMPARATOR in main menu.
Select and confi rm this option to display submenu with direct setting of fl ow levels for changing of output
4.
status (submenu options Qa, Qb), mode indicating active status of the output COMP1(2) (e.g.. Q<Qa) and
hystheresis.
HYST. Complete setting in the same way as described in paragraph B.1.4.